• LOGIN
Repository logo

BORIS Portal

Bern Open Repository and Information System

  • Publication
  • Projects
  • Funding
  • Research Data
  • Organizations
  • Researchers
  • LOGIN
Repository logo
Unibern.ch
  1. Home
  2. Publications
  3. A nonlinear homogenized finite element analysis of the primary stability of the bone–implant interface
 

A nonlinear homogenized finite element analysis of the primary stability of the bone–implant interface

Options
  • Details
BORIS DOI
10.7892/boris.118251
Date of Publication
October 2018
Publication Type
Article
Division/Institute

Institut für chirurgi...

Author
Ovesy, Marziehorcid-logo
Institut für chirurgische Technologien und Biomechanik (ISTB)
Voumard, Benjaminorcid-logo
Institut für chirurgische Technologien und Biomechanik (ISTB)
Zysset, Philippeorcid-logo
Institut für chirurgische Technologien und Biomechanik (ISTB)
Subject(s)

500 - Science::570 - ...

600 - Technology::610...

600 - Technology::620...

Series
Biomechanics and Modeling in Mechanobiology
ISSN or ISBN (if monograph)
1617-7940
Publisher
Springer-Verlag
Language
English
Publisher DOI
10.1007/s10237-018-1038-3
PubMed ID
29858707
Description
Stability of an implant is defined by its ability to undergo physiological loading–unloading cycles without showing excessive tissue damage and micromotions at the interface. Distinction is usually made between the immediate primary stability and the long-term, secondary stability resulting from the biological healing process. The aim of this research is to numerically investigate the effect of initial implantation press-fit, bone yielding, densification and friction at the interface on the primary stability of a simple bone–implant system subjected to loading–unloading cycles. In order to achieve this goal, human trabecular bone was modeled as a continuous, elasto-plastic tissue with damage and densification, which material constants depend on bone volume fraction and fabric. Implantation press-fit related damage in the bone was simulated by expanding the drilled hole to the outer contour of the implant. The bone–implant interface was then modeled with unilateral contact with friction. The implant was modeled as a rigid body and was subjected to increasing off-axis loading cycles. This modeling approach is able to capture the experimentally observed primary stability in terms of initial stiffness, ultimate force and progression of damage. In addition, it is able to quantify the micromotions around the implant relevant for bone healing and osseointegration. In conclusion, the computationally efficient modeling approach used in this study provides a realistic structural response of the bone–implant interface and represents a powerful tool to explore implant design, implantation press-fit and the resulting risk of implant failure under physiological loading.
Handle
https://boris-portal.unibe.ch/handle/20.500.12422/163169
Show full item
File(s)
FileFile TypeFormatSizeLicensePublisher/Copright statementContent
Ovesy2018_Article_ANonlinearHomogenizedFiniteEle.pdftextAdobe PDF1.57 MBpublishedOpen
BORIS Portal
Bern Open Repository and Information System