• LOGIN
    Login with username and password
Repository logo

BORIS Portal

Bern Open Repository and Information System

  • Publications
  • Projects
  • Funding
  • Research Data
  • Organizations
  • Researchers
  • LOGIN
    Login with username and password
Repository logo
Unibern.ch
  1. Home
  2. Publications
  3. Performance of a Novel Handheld Bioelectrical Impedance Device for Assessing Muscle Mass in Older Inpatients.
 

Performance of a Novel Handheld Bioelectrical Impedance Device for Assessing Muscle Mass in Older Inpatients.

Options
  • Details
BORIS DOI
10.48350/169417
Date of Publication
2022
Publication Type
Article
Division/Institute

Clinical Trials Unit ...

Geriatrische Universi...

Contributor
Stuck, Anna
Geriatrische Universitätsklinik
Weber, A
Wittwer, R
Limacher, Andreasorcid-logo
Clinical Trials Unit Bern (CTU)
Kressig, R W
Subject(s)

600 - Technology::610...

Series
Journal of Frailty & Aging
ISSN or ISBN (if monograph)
2273-4309
Publisher
Springer
Language
English
Publisher DOI
10.14283/jfa.2021.40
PubMed ID
35441192
Uncontrolled Keywords

Appendicular skeletal...

Description
OBJECTIVES

To investigate practicality and repeatability of a handheld compared to a state-of-the-art multisegmental bioelectrical impedance analysis (BIA) device to facilitate screening of sarcopenia in older inpatients.

DESIGN AND SETTING

Cross-sectional study in a geriatric rehabilitation hospital.

PARTICIPANTS

207 inpatients aged 70+.

MEASUREMENTS

In a first phase, appendicular skeletal muscle mass index (ASMI) was measured using the handheld Biody xpertZm II BIA device (n=100). In a second phase, ASMI was obtained using the multisegmental Biacorpus RX 4004M device (n=107). Repeatability of BIA devices was compared in subgroups of patients (handheld BIA device: n=36, multisegmental BIA device: n=46) by intra-class correlation (ICC) and Bland-Altman plots.

RESULTS

Overall, measurement failure was seen in 31 patients (31%) tested with the handheld BIA device compared to one patient (0.9%) using the multisegmental BIA device (p<0.001). Main reasons for measurement failure were inability of patients to adopt the position necessary to use the handheld BIA device and device failure. The mean difference of two ASMI measurements in the same patient was 0.32 (sd 0.85) using the handheld BIA device compared to 0.02 kg/m2 (sd 0.07) using the multisegmental device (adjusted mean difference between both groups -0.35, 95% confidence interval (CI) -0.61 to -0.09 kg/m2). Congruently, Bland-Altman plots showed poor agreement with the handheld compared to the multisegmental BIA device.

CONCLUSION

The handheld BIA device is neither a practical nor reliable device for assessing muscle mass in older rehabilitation inpatients.
Handle
https://boris-portal.unibe.ch/handle/20.500.12422/70283
Show full item
File(s)
FileFile TypeFormatSizeLicensePublisher/Copright statementContent
Stuck2022_Article_PerformanceOfANovelHandheldBio.pdftextAdobe PDF659.41 KBAttribution (CC BY 4.0)publishedOpen
BORIS Portal
Bern Open Repository and Information System
Build: 396f6f [24.09. 11:22]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo