  1. Home
  2. Publications
  3. Evaluation of six videolaryngoscopes in 720 patients with a simulated difficult airway: a multicentre randomized controlled trial.
 

Evaluation of six videolaryngoscopes in 720 patients with a simulated difficult airway: a multicentre randomized controlled trial.

Options
  • Details
  • Files
BORIS DOI
10.7892/boris.93958
Publisher DOI
10.1093/bja/aew058
PubMed ID
27106971
Description
BACKGROUND

Videolaryngoscopes are aggressively marketed, but independent evaluation in difficult airways is scarce. This multicentre, prospective randomized controlled trial evaluates six videolaryngoscopes in patients with a simulated difficult airway.

METHODS

With ethics committee approval and written informed consent, 12 senior anaesthetists intubated the trachea of 720 patients. A cervical collar limited mouth opening and neck movement, making intubation difficult. We evaluated three unchannelled (C-MAC™ D-blade, GlideScope™, and McGrath™) and three channelled videolaryngoscopes (Airtraq™, A.P. Advance™ difficult airway blade, and KingVision™). The primary outcome was first-attempt intubation success rate. Secondary outcomes included overall success rate, laryngeal view, intubation times, and side-effects. The primary hypothesis for every videolaryngoscope was that the 95% confidence interval of first-attempt success rate is ≥90%.

RESULTS

Mouth opening was decreased from 46 (sd 7) to 23 (3) mm with the cervical collar. First-attempt success rates were 98% (McGrath™), 95% (C-MAC™ D-blade), 87% (KingVision™), 85% (GlideScope™ and Airtraq™), and 37% (A.P. Advance™, P<0.01). The 95% confidence interval of first-attempt success rate was >90% only for the McGrath™. Overall success, laryngeal view, and intubation times differed significantly between videolaryngoscopes (all P<0.01). Side-effects were minor.

CONCLUSIONS

This trial revealed differences in the performance of six videolaryngoscopes in 720 patients with restricted neck movement and limited mouth opening. In this setting, first-attempt success rates were 85-98%, except for the A.P. Advance™ difficult airway blade. Highest success and lowest tissue trauma rates were achieved by the McGrath™ and C-MAC™ D-blade, highlighting the importance of the videolaryngoscope blade design.

CLINICAL TRIAL REGISTRATION

ClinicalTrials.gov: identifier NCT01692535.
Date of Publication
2016-05
Publication Type
Article
Subject(s)
600 - Technology::610 - Medicine & health
Keyword(s)
anaesthetic techniques
•
laryngoscopy; equipment
•
airway; intubation
•
tracheal tube
Language(s)
en
Contributor(s)
Kleine-Brüggeney, Maren
Universitätsklinik für Anästhesiologie und Schmerztherapie
Greif, Robertorcid-logo
Universitätsklinik für Anästhesiologie und Schmerztherapie
Schoettker, P
Savoldelli, G L
Nabecker, Sabine
Universitätsklinik für Anästhesiologie und Schmerztherapie
Theiler, Lorenz
Universitätsklinik für Anästhesiologie und Schmerztherapie
Additional Credits
Universitätsklinik für Anästhesiologie und Schmerztherapie
Series
British journal of anaesthesia
Publisher
Oxford University Press
ISSN
0007-0912
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: 9f4e9a [ 5.02. 18:48]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo