Cadamuro, JanneJanneCadamuroMrazek, CorneliaCorneliaMrazekLeichtle, Alexander BenediktAlexander BenediktLeichtleKipman, UlrikeUlrikeKipmanFelder, Thomas K.Thomas K.FelderWiedemann, HelmutHelmutWiedemannOberkofler, HannesHannesOberkoflerFiedler, Georg MartinGeorg MartinFiedlerHaschke-Becher, ElisabethElisabethHaschke-Becher2024-10-082024-10-082018https://boris-portal.unibe.ch/handle/20.500.12422/63737Introduction: Although centrifugation is performed in almost every blood sample, recommendations on duration and g-force are heterogeneo-us and mostly based on expert opinions. In order to unify this step in a fully automated laboratory, we aimed to evaluate different centrifugation settings and their influence on the results of routine clinical chemistry analytes. Materials and methods: We collected blood from 41 healthy volunteers into BD Vacutainer PST II-heparin-gel- (LiHepGel), BD Vacutainer SST II-serum-, and BD Vacutainer Barricor heparin-tubes with a mechanical separator (LiHepBar). Tubes were centrifuged at 2000xg for 10 minutes and 3000xg for 7 and 5 minutes, respectively. Subsequently 60 and 21 clinical chemistry analytes were measured in plasma and serum samples, respec-tively, using a Roche COBAS instrument. Results: High sensitive Troponin T, pregnancy-associated plasma protein A, ß human chorionic gonadotropin and rheumatoid factor had to be excluded from statistical evaluation as many of the respective results were below the measuring range. Except of free haemoglobin (fHb) measure-ments, no analyte result was altered by the use of shorter centrifugation times at higher g-forces. Comparing LiHepBar to LiHepGel tubes at different centrifugation setting, we found higher lactate-dehydrogenase (LD) (P = 0.003 to < 0.001) and lower bicarbonate values (P = 0.049 to 0.008) in the lat ter. Conclusions: Serum and heparin samples may be centrifuged at higher speed (3000xg) for a shorter amount of time (5 minutes) without alteration of the analytes tested in this study. When using LiHepBar tubes for blood collection, a separate LD reference value might be needed.en600 - Technology::610 - Medicine & healthInfluence of centrifugation conditions on the results of 77 routine clinical chemistry analytes using standard vacuum blood collection tubes and the new BD-Barricor tubesarticle10.7892/boris.1258212918779710.11613/BM.2018.010704