  3. A comparative study of water perfusion catheters and microtip transducer catheters for urethral pressure measurements
 

A comparative study of water perfusion catheters and microtip transducer catheters for urethral pressure measurements

Options
  • Details
  • Files
BORIS DOI
10.7892/boris.21174
Publisher DOI
10.1007/s00192-006-0255-y
PubMed ID
17131168
Description
The aim of this study was to compare the maximum urethral closure pressure (MUCP) measures with two different techniques: water perfused catheter and microtip transducer catheters with respect to reproducibility and comparability for urethral pressure measurements. Eighteen women with stress urinary incontinence had repeat static urethral pressure profilometry on a different day using a dual microtip transducer and water perfused catheter (Brown and Wickham). The investigators were blinded to the results of the other. The microtip measurements were taken in the 45 degrees upright sitting position with the patient at rest at a bladder capacity of 250 ml using an 8 Fr Gaeltec(R) double microtip transducer withdrawn at 1 mm/s, and the transducer was orientated in the three o'clock position. Three different measures were taken for each patient. Three water perfusion measurements were performed with the patient at rest in the 45 degrees upright position at a bladder capacity of 250 ml using an 8 Fr BARD dual lumen catheter withdrawn at 1 mm/s. The mean water perfusion MUCP measure was 26.1 cm H(2)0, significantly lower than the mean microtip measure of 35.7 cm H(2)0. The correlation coefficient comparing each water perfusion measurement with the other water perfusion measures in the same patient was excellent, at 0.95 (p = 0.01). Correlation coefficient comparing each microtip measure with the other microtip measure in the same patient was also good, ranging from 0.70 to 0.80. This study confirms that both water perfusion catheters and microtip transducers have excellent or very good reproducibility with an acceptable intraindividual variation for both methods.
Date of Publication
2007
Publication Type
article
Language(s)
en
Contributor(s)
Kuhn, Annette
Universitätsklinik für Frauenheilkunde
Nager, Charles W
Hawkins, Emma
Schulz, Jane
Stanton, Stuart L
Additional Credits
Universitätsklinik für Frauenheilkunde
Series
International urogynecology journal
Publisher
Springer-Verlag
ISSN
0937-3462
ISBN
17131168
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: 4f1f0f [ 1.12. 12:07]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo