  3. Virtual Reality-Based and Conventional Visual Field Examination Comparison in Healthy and Glaucoma Patients.
 

Virtual Reality-Based and Conventional Visual Field Examination Comparison in Healthy and Glaucoma Patients.

Options
  • Details
  • Files
BORIS DOI
10.48350/160007
Publisher DOI
10.1167/tvst.10.12.10
PubMed ID
34614166
Description
Purpose

Clinically evaluate the noninferiority of a custom virtual reality (VR) perimetry system when compared to a clinically and routinely used perimeter on both healthy subjects and glaucoma patients.

Methods

We use a custom-designed VR perimetry system tailored for visual field testing. The system uses Oculus Quest VR headset (Facebook Technologies, LLC, Bern, Switzerland), that includes a clicker for participant response feedback. A prospective, single center, study was conducted at the Department of Ophthalmology of the Bern University Hospital (Bern, Switzerland) for 12 months. Of the 114 participants recruited 70 subjects (36 healthy and 34 glaucoma patients with early to moderate visual field loss) were included in the study. Participants underwent perimetry tests on an Octopus 900 (Haag-Streit, Köniz, Switzerland) as well as on the custom VR perimeter. In both cases, standard dynamic strategy (DS) was used in conjunction with the G testing pattern. Collected visual fields (VFs) from both devices were then analyzed and compared.

Results

High mean defect (MD) correlations between the two systems (Spearman, ρ ≥ 0.75) were obtained. The VR system was found to slightly underestimate VF defects in glaucoma subjects (1.4 dB). No significant bias was found with respect to eccentricity or subject age. On average, a similar number of stimuli presentations per VF was necessary when measuring glaucoma patients and healthy subjects.

Conclusions

This study demonstrates that a clinically used perimeter and the proposed VR perimetry system have comparable performances with respect to a number of perimetry parameters in healthy and glaucoma patients with early to moderate visual field loss.

Translational Relevance

This suggests that VR perimeters have the potential to assess VFs with high enough confidence, whereby alleviating challenges in current perimetry practices by providing a portable and more accessible visual field test.
Date of Publication
2021-10-04
Publication Type
Article
Subject(s)
500 - Science::570 - Life sciences; biology
600 - Technology::610 - Medicine & health
Language(s)
en
Contributor(s)
Stapelfeldt, Jan
ARTORG Center for Biomedical Engineering Research
Kucur, Serife Sedaorcid-logo
ARTORG Center - Ophthalmic Technology Lab
ARTORG Center - Artificial Intelligence in Medical Image Computing
Huber, Nina Diana
ARTORG Center for Biomedical Engineering Research
Höhn, René Gerhard Joachimorcid-logo
Universitätsklinik für Augenheilkunde
Sznitman, Raphaelorcid-logo
ARTORG Center - Artificial Intelligence in Medical Image Computing
Additional Credits
ARTORG Center for Biomedical Engineering Research
ARTORG Center - Ophthalmic Technology Lab
Universitätsklinik für Augenheilkunde
ARTORG Center - Artificial Intelligence in Medical Image Computing
Series
Translational vision science & technology
Publisher
Association for Research in Vision and Ophthalmology
ISSN
2164-2591
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: 9f4e9a [ 5.02. 18:48]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo