CT of the medial clavicular epiphysis for forensic age estimation - raised arms position recommended.
Options
BORIS DOI
Publisher DOI
PubMed ID
40439937
Description
Objective
To compare the effect of arm positioning on radiation dose, scan length, and image noise in computed tomography (CT) scans of the medial clavicular epiphysis for forensic age estimation performed with the arms alongside the body (arms-down) versus elevated above the head (arms-up).Methods
Twenty consecutive CT scans were analysed, ten performed with arms-down and ten with arms-up. The scans were conducted at 120 kVp and 37 mAs reference tube current. Scan length extended from 10 mm above to 10 mm below the medial clavicular epiphysis. Dose-relevant parameters (effective CT tube current, volume CT dose index (CTDIvol), CT dose length product (DLP), and effective dose) as well as scan length and image noise were compared between arms-up and arms-down CT scans.Results
Population characteristics: 19 males, 1 female; mean weight 65.8 ± 9.2 kg; height 174.6 ± 7.8 cm; and body mass index (BMI) 21.6 ± 2.5 kg/m². Dose-relevant parameters were significantly lower with arms-up compared to arms-down (effective tube current: 80.9 ± 21.7 mAs vs. 146.0 ± 47.5 mAs, p = 0.001; CTDIvol: 5.5 ± 1.5 mGy vs. 9.9 ± 3.2 mGy, p = 0.001; DLP: 40.2 ± 13.7 mGy*cm vs. 63.8 ± 21.9 mGy*cm, p = 0.010; effective dose: 0.6 ± 0.2 mSv vs. 0.9 ± 0.3 mSv, p = 0.010). No significant differences were found in scan length, image noise, or population characteristics.Conclusions
Removing the arms from the CT beam path reduced radiation dose by 33% without affecting scan length or image noise. Given the importance of dose optimisation in non-medical examinations of potentially minor individuals, CT scans of the medial clavicular epiphysis should be performed with arms elevated above the head.
To compare the effect of arm positioning on radiation dose, scan length, and image noise in computed tomography (CT) scans of the medial clavicular epiphysis for forensic age estimation performed with the arms alongside the body (arms-down) versus elevated above the head (arms-up).Methods
Twenty consecutive CT scans were analysed, ten performed with arms-down and ten with arms-up. The scans were conducted at 120 kVp and 37 mAs reference tube current. Scan length extended from 10 mm above to 10 mm below the medial clavicular epiphysis. Dose-relevant parameters (effective CT tube current, volume CT dose index (CTDIvol), CT dose length product (DLP), and effective dose) as well as scan length and image noise were compared between arms-up and arms-down CT scans.Results
Population characteristics: 19 males, 1 female; mean weight 65.8 ± 9.2 kg; height 174.6 ± 7.8 cm; and body mass index (BMI) 21.6 ± 2.5 kg/m². Dose-relevant parameters were significantly lower with arms-up compared to arms-down (effective tube current: 80.9 ± 21.7 mAs vs. 146.0 ± 47.5 mAs, p = 0.001; CTDIvol: 5.5 ± 1.5 mGy vs. 9.9 ± 3.2 mGy, p = 0.001; DLP: 40.2 ± 13.7 mGy*cm vs. 63.8 ± 21.9 mGy*cm, p = 0.010; effective dose: 0.6 ± 0.2 mSv vs. 0.9 ± 0.3 mSv, p = 0.010). No significant differences were found in scan length, image noise, or population characteristics.Conclusions
Removing the arms from the CT beam path reduced radiation dose by 33% without affecting scan length or image noise. Given the importance of dose optimisation in non-medical examinations of potentially minor individuals, CT scans of the medial clavicular epiphysis should be performed with arms elevated above the head.
Date of Publication
2025-09
Publication Type
Article
Keyword(s)
Arm positioning
•
Computed tomography
•
Forensic age estimation
•
Medial clavicular epiphysis
•
Radiation protection
Language(s)
en
Contributor(s)
Kuhnen, Saskia C | |
Schmeling, Andreas | |
Ith, Michael |
Series
International Journal of Legal Medicine
Publisher
Springer
ISSN
1437-1596
0937-9827
Access(Rights)
open.access