  2. Publications
  3. Comparison of a Novel Temperature-Controlled Diamond-Tip Catheter and a Power-Controlled Gold-Tip Catheter for the Irrigated Ablation of Cavotricuspid Isthmus-Dependent Atrial Flutter.
 

Comparison of a Novel Temperature-Controlled Diamond-Tip Catheter and a Power-Controlled Gold-Tip Catheter for the Irrigated Ablation of Cavotricuspid Isthmus-Dependent Atrial Flutter.

Options
  • Details
  • Files
BORIS DOI
10.48620/85703
Publisher DOI
10.3390/jcm14030701
PubMed ID
39941371
Description
Background/Objectives: Radiofrequency (RF) ablation of the cavotricuspid isthmus (CTI) is a recommended treatment option for typical atrial flutter (AFL). While power-controlled ablation has been the current standard, a novel temperature-controlled ablation system has been introduced. We aimed to compare the procedural efficacy and one-year outcome of a temperature-controlled diamond-tip catheter with an established power-controlled gold-tip catheter. Methods: Consecutive patients undergoing ablation of CTI-dependent AFL using a power-controlled catheter or the novel temperature-controlled catheter were enrolled. Patients were followed up using a 7-day electrocardiogram after 3, 6, and 12 months. The primary endpoint was acute efficacy (procedural success, total RF, and procedure time). The secondary endpoint was the recurrence of typical AFL during follow-up. Results: In total, 38 patients undergoing temperature-controlled ablation were enrolled and compared to 283 patients undergoing power-controlled ablation. A bidirectional CTI block was achieved in 100% in the temperature-controlled group and 97.5% in the power-controlled group (p = 0.7). The total RF time (median: 192 sec (IQR 138-311) vs. 643 sec (IQR 386-1079), p < 0.001) and total procedure time (median: 45 min (IQR 34-57) vs. 52 min (IQR 39-70), p = 0.01) were shorter with temperature-controlled ablation. At the one-year follow-up, there was no difference in the recurrence of typical AFL between groups. Conclusions: Utilization of temperature-controlled ablation for typical AFL increased procedural efficiency with shorter RF and procedure times compared to power-controlled ablation. The recurrence rate of typical AFL after one year was low and did not differ amongst groups.
Date of Publication
2025-01-22
Publication Type
Article
Subject(s)
600 - Technology::610 - Medicine & health
Keyword(s)
ablation
•
atrial flutter
•
catheter
•
cavotricuspid isthmus
Language(s)
en
Contributor(s)
Auf der Maur, Eric Jan
Institute of Tissue Medicine and Pathology, Autopsy
Kueffer, Thomasorcid-logo
Clinic of Cardiology
Thalmann, Gregor
Clinic of Cardiology
Kozhuharov, Nikola A.
Clinic of Cardiology
Galuszka, Oskar
Clinic of Cardiology
Iqbal, Salik Ur Rehman
Clinic of Cardiology
Roten, Laurentorcid-logo
Clinic of Cardiology
Madaffari, Antonio
Reichlin, Tobias
Clinic of Cardiology - Partial Clinic
Noti, Fabian
Clinic of Cardiology
Tanner, Hildegard
Clinic of Cardiology
Haeberlin, Andreasorcid-logo
Clinic of Cardiology
Servatius, Helge
Clinic of Cardiology
Additional Credits
Clinic of Cardiology
Institute of Tissue Medicine and Pathology, Autopsy
Clinic of Cardiology - Partial Clinic
Series
Journal of Clinical Medicine
Publisher
MDPI
ISSN
2077-0383
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: ae9592 [15.12. 16:43]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo