• LOGIN
Repository logo

BORIS Portal

Bern Open Repository and Information System

  • Publication
  • Projects
  • Funding
  • Research Data
  • Organizations
  • Researchers
  • LOGIN
Repository logo
Unibern.ch
  1. Home
  2. Publications
  3. Fuzzy logic control of mechanical ventilation during anaesthesia.
 

Fuzzy logic control of mechanical ventilation during anaesthesia.

Options
  • Details
BORIS DOI
10.7892/boris.115155
Date of Publication
1996
Publication Type
Article
Division/Institute

Universitätsklinik fü...

Author
Schäublin, J.
Derighetti, M.
Feigenwinter, P.
Petersen, Steen
Universitätsklinik für Anästhesiologie und Schmerztherapie
Zbinden, Alex Martin
Universitätsklinik für Anästhesiologie und Schmerztherapie
Subject(s)

600 - Technology::610...

Series
British journal of anaesthesia
ISSN or ISBN (if monograph)
0007-0912
Publisher
Oxford University Press
Language
English
Publisher DOI
10.1093/bja/77.5.636
Description
We have examined a new approach, using fuzzy logic, to the closed-loop feedback control of mechanical ventilation during general anaesthesia. This control system automatically adjusts ventilatory frequency (f) and tidal volume (VT) in order to achieve and maintain the end-tidal carbon dioxide fraction (FE'CO2) at a desired level (set-point). The controller attempts to minimize the deviation of both f and VT per kg body weight from 10 bpm and 10 ml kg-1, respectively, and to maintain the plateau airway pressure within suitable limits. In 30 patients, undergoing various surgical procedures, the fuzzy control mode was compared with human ventilation control. For a set-point of FE'CO2 = 4.5 vol% and during measurement periods of 20 min, accuracy, stability and breathing pattern did not differ significantly between fuzzy logic and manual ventilation control. After step-changes in the set-point of FE'CO2 from 4.5 to 5.5 vol% and vice versa, overshoot and rise time did not differ significantly between the two control modes. We conclude that to achieve and maintain a desired FE'CO2 during routine anaesthesia, fuzzy logic feedback control of mechanical ventilation is a reliable and safe mode of control.
Handle
https://boris-portal.unibe.ch/handle/20.500.12422/160941
Show full item
File(s)
FileFile TypeFormatSizeLicensePublisher/Copright statementContent
8957981.pdftextAdobe PDF162.52 KBpublisherpublishedOpen
BORIS Portal
Bern Open Repository and Information System
Build: 360c85 [14.04. 8:05]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo