Noninvasive Quantification of Retinal Microglia Using Widefield Autofluorescence Imaging.
Options
BORIS DOI
Publisher DOI
PubMed ID
28395300
Description
Purpose
To validate widefield autofluorescence (AF) in vivo imaging of the retina in mice expressing green fluorescent protein (gfp) in microglia, and to monitor retinal microglia reconstitution in vivo after lethal irradiation and bone marrow transplantation.
Methods
Transgenic Cx3cr1gfp/gfp and wildtype Balb/c mice were used in this study. A confocal scanning laser ophthalmoscope was used for AF imaging with a 55° and a widefield 102° lens. Intrasession reproducibility was assessed for each lens. To investigate reconstitution in vivo, bone marrow from Cx3cr1gfp/gfp mice was used to rescue lethally irradiated wildtype mice. Data were compared to confocal microscopy of retinal flat mounts.
Results
Both the 55° and the 102° lens produced high resolution images of retinal microglia with similar microglia density. However, compared to the 55° lens, the widefield 102° lens captured approximately 3.6 times more microglia cells (1515 ± 123 cells versus 445 ± 76 cells [mean ± SD], for 102° and 55°, respectively, P < 0.001). No statistical difference in the number of gfp positive cells within corresponding areas was observed within the same imaging session. Imaging of microglia reconstitution showed a similar time course compared to flat mount preparations with an excellent correlation between microglia cell numbers in AF and gfp-stained flat mounts (R = 0.92, P < 0.0001).
Conclusions
Widefield AF imaging of mice with gfp expressing microglia can be used to quantify retinal microglia. In vivo microglia counts corresponded very well with ex vivo counts on retinal flat mounts. As such, AF imaging can largely replace ex vivo quantification.
To validate widefield autofluorescence (AF) in vivo imaging of the retina in mice expressing green fluorescent protein (gfp) in microglia, and to monitor retinal microglia reconstitution in vivo after lethal irradiation and bone marrow transplantation.
Methods
Transgenic Cx3cr1gfp/gfp and wildtype Balb/c mice were used in this study. A confocal scanning laser ophthalmoscope was used for AF imaging with a 55° and a widefield 102° lens. Intrasession reproducibility was assessed for each lens. To investigate reconstitution in vivo, bone marrow from Cx3cr1gfp/gfp mice was used to rescue lethally irradiated wildtype mice. Data were compared to confocal microscopy of retinal flat mounts.
Results
Both the 55° and the 102° lens produced high resolution images of retinal microglia with similar microglia density. However, compared to the 55° lens, the widefield 102° lens captured approximately 3.6 times more microglia cells (1515 ± 123 cells versus 445 ± 76 cells [mean ± SD], for 102° and 55°, respectively, P < 0.001). No statistical difference in the number of gfp positive cells within corresponding areas was observed within the same imaging session. Imaging of microglia reconstitution showed a similar time course compared to flat mount preparations with an excellent correlation between microglia cell numbers in AF and gfp-stained flat mounts (R = 0.92, P < 0.0001).
Conclusions
Widefield AF imaging of mice with gfp expressing microglia can be used to quantify retinal microglia. In vivo microglia counts corresponded very well with ex vivo counts on retinal flat mounts. As such, AF imaging can largely replace ex vivo quantification.
Date of Publication
2017-04-01
Publication Type
Article
Subject(s)
Language(s)
en
Contributor(s)
Kokona, Despina | |
Schneider, Nadia | |
Giannakaki-Zimmermann, Helena | |
Jovanovic, Joel |
Additional Credits
Series
Investigative ophthalmology & visual science
Publisher
Association for Research in Vision and Ophthalmology
ISSN
0146-0404
Access(Rights)
open.access