Reemergence of Mycobacterium chimaera in Heater-Cooler Units despite Intensified Cleaning and Disinfection Protocol.
Options
BORIS DOI
Publisher DOI
PubMed ID
27649345
Description
Invasive Mycobacterium chimaera infections after open-heart surgery have been reported internationally. These devastating infections result from aerosols generated by contaminated heater-cooler units used with extracorporeal circulation during surgery. Despite intensified cleaning and disinfection, surveillance samples from factory-new units acquired during 2014 grew nontuberculous mycobacteria after a median of 174 days.
Date of Publication
2016-10
Publication Type
Article
Subject(s)
600 - Technology::610 - Medicine & health
Keyword(s)
Mycobacterium chimaera
•
bacteria
•
disinfection
•
heater–cooler units
•
hospital equipment
•
hospital infections
•
nontuberculous mycobacterial infections
•
nosocomial infections
•
tuberculosis and other mycobacteria
Language(s)
en
Contributor(s)
Schreiber, Peter W | |
Kuster, Stefan P | |
Hasse, Barbara | |
Bayard, Cornelia | |
Rüegg, Christian | |
Kohler, Philipp | |
Keller, Peter M | |
Bloemberg, Guido V | |
Maisano, Francesco | |
Bettex, Dominique | |
Halbe, Maximilian | |
Sax, Hugo |
Additional Credits
Universitätsklinik für Infektiologie
Series
Emerging infectious diseases
Publisher
U.S. National Center for Infectious Diseases
ISSN
1080-6040
Access(Rights)
open.access