Description
Phalangeal fractures are common, particularly in younger patients, leading to a large economic burden due to higher incident rates among patients of working age. In traumatic cases where the fracture may be unstable, plate fixation has grown in popularity due to its greater construct rigidity. However, these metal plates have increased reoperation rates due to inflammation of the surrounding soft tissue. To overcome these challenges, a novel osteosynthesis platform, AdhFix, has been developed. This method uses a light-curable polymer that can be shaped around traditional metal screws to create a plate-like structure that has been shown to not induce soft tissue adhesions. However, to effectively evaluate any novel osteosynthesis device, the biomechanical environment must first be understood. In this study, the internal loads in a phalangeal plate osteosynthesis were measured under simulated rehabilitation exercises. In a human hand cadaver study, a plastic plate with known biomechanical properties was used to fix a 3 mm osteotomy and each finger was fully flexed to mimic traditional rehabilitation exercises. The displacements of the bone fragments were tracked with a stereographic camera system and coupled with specimen specific finite element (FE) models to calculate the internal loads in the osteosynthesis. Following this, AdhFix patches were created and monotonically tested under similar conditions to determine survival of the novel technique. The internal bending moment in the osteosynthesis was 6.78 ± 1.62 Nmm and none of the AdhFix patches failed under the monotonic rehabilitation exercises. This study demonstrates a method to calculate the internal loads on an osteosynthesis device during non-load bearing exercises and that the novel AdhFix solution did not fail under traditional rehabilitation protocols in this controlled setting. Further studies are required prior to clinical application.
Date of Publication
2024
Publication Type
Article
Keyword(s)
CT derived models
•
customizable osteosythesis
•
finite element modeling
•
non-contact measurement
•
patient-specific treatment
Language(s)
en
Contributor(s)
Schwarzenberg, Peter
Colding-Rasmussen, Thomas
Hutchinson, Daniel J
San Jacinto Garcia, Jorge
Granskog, Viktor
Mørk Petersen, Michael
Pastor, Tatjana
Clinic of Plastic and Hand Surgery
Weis, Tine
Malkoch, Michael
Nai En Tierp-Wong, Christian
Varga, Peter
Additional Credits
Clinic of Plastic and Hand Surgery
Series
Frontiers in Bioengineering and Biotechnology
Publisher
Frontiers Media
ISSN
2296-4185
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: 9f4e9a [ 5.02. 18:48]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo