Publication:
A Swiss cheese error detection method for real-time EPID-based quality assurance and error prevention.

cris.virtual.author-orcid0000-0002-3911-3403
cris.virtualsource.author-orcidfb089609-c18a-4ab4-bf39-a2d95248a9a9
cris.virtualsource.author-orcidc171715f-9c9f-4166-9093-7f89b4d6ffe9
dc.contributor.authorPassarge, Michelle
dc.contributor.authorFix, Michael
dc.contributor.authorManser, Peter
dc.contributor.authorStampanoni, Marco F M
dc.contributor.authorSiebers, Jeffrey V
dc.date.accessioned2024-10-25T05:05:46Z
dc.date.available2024-10-25T05:05:46Z
dc.date.issued2017-04
dc.description.abstractPURPOSE To develop a robust and efficient process that detects relevant dose errors (dose errors of ≥5%) in external beam radiation therapy and directly indicates the origin of the error. The process is illustrated in the context of electronic portal imaging device (EPID)-based angle-resolved volumetric modulated arc therapy (VMAT) quality assurance (QA), particularly as would be implemented in a real-time monitoring program. METHODS A Swiss cheese error detection (SCED) method was created as a paradigm for a cine EPID-based during-treatment QA. For VMAT, the method compares a treatment-plan-based reference set of EPID images with images acquired over each 2° gantry angle interval. The process utilizes a sequence of independent consecutively executed error detection tests: an aperture check that verifies infield radiation delivery and ensures no out-of-field radiation; output normalization checks at two different stages; global image alignment check to examine if rotation, scaling and translation are within tolerances; pixel intensity check containing the standard gamma evaluation (3%, 3 mm) and pixel intensity deviation checks including and excluding high dose gradient regions. Tolerances for each check were determined. To test the SCED method, 12 different types of errors were selected to modify the original plan. A series of angle-resolved predicted EPID images was artificially generated for each test case, resulting in a sequence of pre-calculated frames for each modified treatment plan. The SCED method was applied multiple times for each test case to assess the ability to detect introduced plan variations. To compare the performance of the SCED process with that of a standard gamma analysis, both error detection methods were applied to the generated test cases with realistic noise variations. RESULTS Averaged over ten test runs, 95.1% of all plan variations that resulted in relevant patient dose errors were detected within 2° and 100% within 14° (<4% of patient dose delivery). Including cases that led to slightly modified but clinically equivalent plans, 89.1% were detected by the SCED method within 2°. Based on the type of check that detected the error, determination of error sources was achieved. With noise ranging from no random noise to four times the established noise value, the averaged relevant dose error detection rate of the SCED method was between 94.0% and 95.8% and that of gamma between 82.8% and 89.8%. CONCLUSIONS An EPID-frame-based error detection process for VMAT deliveries was successfully designed and tested via simulations. The SCED method was inspected for robustness with realistic noise variations, demonstrating that it has the potential to detect a large majority of relevant dose errors. Compared to a typical (3%, 3 mm) gamma analysis, the SCED method produced a higher detection rate for all introduced dose errors, identified errors in an earlier stage, displayed a higher robustness to noise variations and indicated the error source. This article is protected by copyright. All rights reserved.
dc.description.numberOfPages12
dc.description.sponsorshipUniversitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
dc.identifier.doi10.7892/boris.95191
dc.identifier.pmid28134989
dc.identifier.publisherDOI10.1002/mp.12142
dc.identifier.urihttps://boris-portal.unibe.ch/handle/20.500.12422/149662
dc.language.isoen
dc.publisherAmerican Association of Physicists in Medicine AAPM
dc.relation.ispartofMedical physics
dc.relation.issn0094-2405
dc.relation.organizationDCD5A442BAE6E17DE0405C82790C4DE2
dc.subjectEPID VMAT error detection real-time quality assurance treatment verification
dc.subject.ddc600 - Technology::610 - Medicine & health
dc.titleA Swiss cheese error detection method for real-time EPID-based quality assurance and error prevention.
dc.typearticle
dspace.entity.typePublication
dspace.file.typetext
oaire.citation.endPage1223
oaire.citation.issue4
oaire.citation.startPage1212
oaire.citation.volume44
oairecerif.author.affiliationUniversitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
oairecerif.author.affiliationUniversitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
unibe.contributor.rolecreator
unibe.contributor.rolecreator
unibe.contributor.rolecreator
unibe.contributor.rolecreator
unibe.contributor.rolecreator
unibe.date.licenseChanged2019-10-24 18:01:50
unibe.description.ispublishedpub
unibe.eprints.legacyId95191
unibe.journal.abbrevTitleMED PHYS
unibe.refereedTRUE
unibe.subtype.articlejournal

Files

Original bundle
Now showing 1 - 1 of 1
Name:
Passarge_et_al-2017-Medical_Physics.pdf
Size:
341.91 KB
Format:
Adobe Portable Document Format
File Type:
text
License:
publisher
Content:
published

Collections