Multichannel Esophageal Heart Rate Monitoring of Preterm Infants
Options
BORIS DOI
Publisher DOI
PubMed ID
33044926
Description
Objective: Autonomic dysregulation in preterm infants requires continuous monitoring of vital signs such as heart rate over days to months. Unfortunately, common surface electrodes are prone to electrocardiography (ECG) signal artifacts and cause serious skin irritations during long-term use. In contrast, esophageal ECG is known to be very sensitive due to the proximity of electrodes and heart and insensitive to external influences. This study addresses if multichannel esophageal ECG qualifies for heart rate monitoring in preterm infants.
Methods: We recorded esophageal leads with a multi-electrode gastric feeding tube in a clinical study with 13 neonates and compared the heartbeat detection performance with standard surface leads. A computationally simple and versatile ECG wave detection algorithm was used.
Results: Multichannel esophageal ECG manifested heartbeat sensitivity and positive predictive value greater than 98.5% and significant less false negative (FN) ECG waves as compared to surface ECG due to site-typical electrode motion artifacts. False positive bradycardia as indicated with more than 13 consecutive FN ECG waves was equally expectable in esophageal and surface channels. No adverse events were reported for the multi-electrode gastric feeding tube.
Conclusion: Heart rate monitoring of preterm infants with multiple esophageal electrodes is considered as feasible and reliable. Less signal artifacts will improve the detection of bradycardia, which is crucial for immediate interventions, and reduce alarm fatigue.
Significance: Due to the possibility to integrate the multichannel ECG into a gastric feeding tube and meanwhile omit harmful skin electrodes, the presented system has great potential to facilitate future intensive care of preterm infants.
Methods: We recorded esophageal leads with a multi-electrode gastric feeding tube in a clinical study with 13 neonates and compared the heartbeat detection performance with standard surface leads. A computationally simple and versatile ECG wave detection algorithm was used.
Results: Multichannel esophageal ECG manifested heartbeat sensitivity and positive predictive value greater than 98.5% and significant less false negative (FN) ECG waves as compared to surface ECG due to site-typical electrode motion artifacts. False positive bradycardia as indicated with more than 13 consecutive FN ECG waves was equally expectable in esophageal and surface channels. No adverse events were reported for the multi-electrode gastric feeding tube.
Conclusion: Heart rate monitoring of preterm infants with multiple esophageal electrodes is considered as feasible and reliable. Less signal artifacts will improve the detection of bradycardia, which is crucial for immediate interventions, and reduce alarm fatigue.
Significance: Due to the possibility to integrate the multichannel ECG into a gastric feeding tube and meanwhile omit harmful skin electrodes, the presented system has great potential to facilitate future intensive care of preterm infants.
Date of Publication
2021-05-21
Publication Type
Article
Subject(s)
Language(s)
en
Contributor(s)
Simmen, Patrizia | |
Kreuzer, Samuel | |
Thomet, Manuel | |
Suter, Lilian | |
Jesacher, Barbara | |
Tran, Phuong-Anh | |
Schulzke, Sven | |
Jost, Kerstin | |
Niederhauser, Thomas |
Additional Credits
Series
IEEE transactions on biomedical engineering
Publisher
Institute of Electrical and Electronics Engineers IEEE
ISSN
0018-9294
Access(Rights)
restricted