  3. NEMA NU 2-2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET/CT.
 

NEMA NU 2-2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET/CT.

Options
  • Details
  • Files
BORIS DOI
10.48350/167400
Publisher DOI
10.1007/s00259-022-05751-7
PubMed ID
35284970
Description
PURPOSE

The DMI PET/CT is a modular silicon photomultiplier-based scanner with an axial field-of-view (FOV) between 15 and 25 cm depending on ring configuration (3, 4, or 5 rings). A new generation of the system includes a reengineered detector module, featuring improved electronics and an additional 6th ring, extending the axial FOV to 30 cm. We report on the performance evaluation of the 6-ring upgraded Generation 2 (Gen2) system while values are also reported for the 5-ring configuration of the very same system prior to the upgrade.

METHODS

PET performance was evaluated using the NEMA NU 2-2018 standard for spatial resolution, sensitivity, image quality, count rate performance, timing resolution, and image co-registration accuracy. Patient images were used to assess image quality.

RESULTS

The average system sensitivity was measured at 32.76 cps/kBq (~ 47% increase to 5 rings at 22.29 cps/kBq) while noise equivalent count rate peaked at 434.3 kcps corresponding to 23.6 kBq/mL (~ 60% increase to Generation 1 (Gen1) and 39% to Gen2 5 rings). Contrast recovery ranged between 54.5 and 85.8% similar to 5 rings, while the 6 rings provided lower background variability (2.3-8.5% for 5 rings vs 1.9-6.8% for 6 rings) and lower lung error (4.0% for the 5 rings and 3.16% for the 6 rings). Transverse/axial full width at half-maximum (FWHM) at 1 cm (3.79/4.26 mm) and 10 cm (4.29/4.55 mm), scatter fraction (40.2%), energy resolution (9.63%), and time-of-flight (TOF) resolution (389.6 ps at 0 kBq/mL) were in line to previously reported values measured across different system configurations. Improved patient image quality is obtained with the 6 rings compared to the 5 rings, while image quality is retained even at reduced scan times, enabling WB dynamic acquisitions.

CONCLUSIONS

The higher sensitivity of the 6-ring DMI compared to the 5-ring configuration may lead to improved image quality of clinical images at reduced scan time. Additionally, it could equally be used to allow improved temporal sampling and/or reduced overall scan time in dynamic acquisitions. Conversely, temporal sampling and scan time could be traded per application to further drive injected dose at lower levels.
Date of Publication
2022-07
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Keyword(s)
6-Ring Discovery MI 6-ring scanner NEMA NU 2 PET/CT Performance evaluation
Language(s)
en
Contributor(s)
Zeimpekis, Konstantinos
Universitätsklinik für Nuklearmedizin
Kotasidis, Fotis A
Huellner, Martin
Nemirovsky, Alexandra
Kaufmann, Philipp A
Treyer, Valerie
Additional Credits
Universitätsklinik für Nuklearmedizin
Series
European journal of nuclear medicine and molecular imaging
Publisher
Springer
ISSN
1619-7089
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: dd892c [ 9.04. 8:30]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo