Comparison of freehand-navigated and aiming device-navigated targeting of liver lesions
Options
BORIS DOI
Publisher DOI
PubMed ID
23832927
Description
BACKGROUND
Accurate needle placement is crucial for the success of percutaneous radiological needle interventions. We compared three guiding methods using an optical-based navigation system: freehand, using a stereotactic aiming device and active depth control, and using a stereotactic aiming device and passive depth control.
METHODS
For each method, 25 punctures were performed on a non-rigid phantom. Five 1 mm metal screws were used as targets. Time requirements were recorded, and target positioning errors (TPE) were measured on control scans as the distance between needle tip and target.
RESULTS
Time requirements were reduced using the aiming device and passive depth control. The Euclidian TPE was similar for each method (4.6 ± 1.2-4.9 ± 1.7 mm). However, the lateral component was significantly lower when an aiming device was used (2.3 ± 1.3-2.8 ± 1.6 mm with an aiming device vs 4.2 ± 2.0 mm without).
DISCUSSION
Using an aiming device may increase the lateral accuracy of navigated needle insertion.
Accurate needle placement is crucial for the success of percutaneous radiological needle interventions. We compared three guiding methods using an optical-based navigation system: freehand, using a stereotactic aiming device and active depth control, and using a stereotactic aiming device and passive depth control.
METHODS
For each method, 25 punctures were performed on a non-rigid phantom. Five 1 mm metal screws were used as targets. Time requirements were recorded, and target positioning errors (TPE) were measured on control scans as the distance between needle tip and target.
RESULTS
Time requirements were reduced using the aiming device and passive depth control. The Euclidian TPE was similar for each method (4.6 ± 1.2-4.9 ± 1.7 mm). However, the lateral component was significantly lower when an aiming device was used (2.3 ± 1.3-2.8 ± 1.6 mm with an aiming device vs 4.2 ± 2.0 mm without).
DISCUSSION
Using an aiming device may increase the lateral accuracy of navigated needle insertion.
Date of Publication
2014-03
Publication Type
Article
Subject(s)
Keyword(s)
image-guided therapy interventional radiology stereotaxy
Language(s)
en
Additional Credits
Series
International journal of medical robotics and computer assisted surgery
Publisher
Wiley
ISSN
1478-5951
Access(Rights)
restricted