Unibern.ch
  1. Home
  2. Publications
  3. The Freedom SOLO bovine pericardial stentless valve: Single-center experience, outcome and long-term durability
 

The Freedom SOLO bovine pericardial stentless valve: Single-center experience, outcome and long-term durability

Options
  • Details
  • Files
BORIS DOI
10.7892/boris.63202
Publisher DOI
10.1016/j.jtcvs.2015.01.060
PubMed ID
25769775
Description
Abstract

Objectives

We report our institutional experience and long-term results with the Sorin Freedom SOLO bovine pericardial stentless bioprosthesis.

Methods

Between January 2005 and November 2009, 149 patients (mean age 73.6±8.7 years, 68 [45.6%] female) underwent isolated (n=75) or combined (n=74) aortic valve replacement (AVR) using the SOLO in our institution. Follow-up was 100% complete with an average follow-up time of 5.9±2.6 years (maximum 9.6 years) and a total of 885.3 patient years.

Results

Operative (30-day) mortality was 2.7% (1.3% for isolated AVR [n=1] and 4.0% for combined procedures [n=3]). All causes of death were not valve-related. Preoperative peak (mean) gradients of 74.2±23.0 mmHg (48.6 ± 16.3 mmHg) decreased to 15.6±5.4 (8.8±3.0) after AVR, and remained low for up to 9 years. The postoperative effective orifice area (EOA) was 1.6 ±0.57 cm2, 1.90±0.45 cm2, 2.12±0.48 cm2 and 2.20±0.66 cm2 for the valve sizes 21, 23, 25 and 27, respectively; with absence of severe prosthesis-patient-mismatch (PPM) and 0.7% (n=1) moderate PPM. During follow-up, Twenty-six patients experienced structural valve deterioration (SVD) and 14 patients underwent explantation. Kaplan-Meier estimates for freedom from death, explantation and SVD at 9 years averaged 0.57 [0.47‒0.66], 0.82 [0.69‒0.90] and 0.70 [0.57‒0.79], respectively.

Conclusions

The Freedom SOLO stentless aortic valve is safe to implant and shows excellent early and mid-term hemodynamic performance. However, SVD was observed in a substantial number of patients after only 5 ̶ 6 years and the need for explantation increased markedly, suggesting low durability.
Date of Publication
2015-02-10
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Language(s)
en
Contributor(s)
Stanger, Olaf
Universitätsklinik für Herz- und Gefässchirurgie
Bleuel, Irina
Gisler, Fabian
Universitätsklinik für Herz- und Gefässchirurgie
Göber, Volkhard
Universitätsklinik für Herz- und Gefässchirurgie
Reineke, Sylvia
Universitätsklinik für Herz- und Gefässchirurgie
Gahl, Brigitta
Universitätsklinik für Herz- und Gefässchirurgie
Aymard, Thierry
Englberger, Lars
Universitätsklinik für Herz- und Gefässchirurgie
Carrel, Thierry
Universitätsklinik für Herz- und Gefässchirurgie
Tevaearai, Hendrik
Universitätsklinik für Herz- und Gefässchirurgie
Additional Credits
Universitätsklinik für Herz- und Gefässchirurgie
Series
Journal of thoracic and cardiovascular surgery
Publisher
Mosby
ISSN
0022-5223
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: dd892c [ 9.04. 8:30]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo