Performance of a new laser fluorescence device for the detection of occlusal caries in vitro
Options
Publisher DOI
PubMed ID
16431009
Description
The new device DIAGNOdent pen based on red laser light induced fluorescence was introduced for the detection of approximal and occlusal caries. The aim of this study was to test its performance on occlusal surfaces. The new device comes with two different sapphire fibre tips: a cylindrical tip and a conical tip. The two new sapphire fibre tips were used and compared with the tip currently available with DIAGNOdent (DD). METHODS: The teeth were selected from a pool of extracted permanent human molars, which were stored frozen at -20 degrees C, until use. Prior to being measured the teeth were defrosted and cleaned. One hundred and nineteen teeth were selected and measured with the old tip and with the two new tips of the new device by two independent assessments. The teeth were histologically prepared and assessed for caries extension. RESULTS: Specificity values for D(1), D(2) and D(3) ranged between 0.69 and 0.89, sensitivity between 0.78 and 0.96. There were no statistically significant differences obtained between the two tips of the new and the one tip of the old device (p>0.05). Intra-examiner reliability with kappa values of >0.83 was high. CONCLUSIONS: In this study, the new laser fluorescence device performed on occlusal surfaces as well as the available device.
Date of Publication
2006
Publication Type
Article
Language(s)
en
Contributor(s)
Additional Credits
Series
Journal of dentistry
Publisher
Elsevier Science
ISSN
0300-5712
Access(Rights)
metadata.only