• LOGIN
    Login with username and password
Repository logo

BORIS Portal

Bern Open Repository and Information System

  • Publications
  • Projects
  • Funding
  • Research Data
  • Organizations
  • Researchers
  • LOGIN
    Login with username and password
Repository logo
Unibern.ch
  1. Home
  2. Publications
  3. Improved Localization of Implanted Subdural Electrode Contacts on MRI Using an Elastic Image Fusion Algorithm in Invasive EEG Recording
 

Improved Localization of Implanted Subdural Electrode Contacts on MRI Using an Elastic Image Fusion Algorithm in Invasive EEG Recording

Options
  • Details
BORIS DOI
10.7892/boris.58185
Date of Publication
June 23, 2014
Publication Type
Article
Division/Institute

Universitätsklinik fü...

Universitätsklinik fü...

Universitätsinstitut ...

Contributor
Stieglitz, Lennart
Universitätsklinik für Neurochirurgie
Ayer, Christian
Schindler, Kaspar
Universitätsklinik für Neurologie
Oertel, Markus Florian
Universitätsklinik für Neurochirurgie
Wiest, Roland Gerhard Rudi
Universitätsinstitut für Diagnostische und Interventionelle Neuroradiologie
Pollo, Claudio
Universitätsklinik für Neurochirurgie
Subject(s)

600 - Technology::610...

Series
Neurosurgery
ISSN or ISBN (if monograph)
0148-396X
Publisher
Lippincott Williams & Wilkins
Language
English
Publisher DOI
10.1227/NEU.0000000000000473
PubMed ID
24978648
Description
BACKGROUND:

Accurate projection of implanted subdural electrode contacts in presurgical evaluation of pharmacoresistant epilepsy cases by invasive EEG is highly relevant. Linear fusion of CT and MRI images may display the contacts in the wrong position due to brain shift effects.

OBJECTIVE:

A retrospective study in five patients with pharmacoresistant epilepsy was performed to evaluate whether an elastic image fusion algorithm can provide a more accurate projection of the electrode contacts on the pre-implantation MRI as compared to linear fusion.

METHODS:

An automated elastic image fusion algorithm (AEF), a guided elastic image fusion algorithm (GEF), and a standard linear fusion algorithm (LF) were used on preoperative MRI and post-implantation CT scans. Vertical correction of virtual contact positions, total virtual contact shift, corrections of midline shift and brain shifts due to pneumencephalus were measured.

RESULTS:

Both AEF and GEF worked well with all 5 cases. An average midline shift of 1.7mm (SD 1.25) was corrected to 0.4mm (SD 0.8) after AEF and to 0.0mm (SD 0) after GEF. Median virtual distances between contacts and cortical surface were corrected by a significant amount, from 2.3mm after LF to 0.0mm after AEF and GEF (p<.001). Mean total relative corrections of 3.1 mm (SD 1.85) after AEF and 3.0mm (SD 1.77) after GEF were achieved. The tested version of GEF did not achieve a satisfying virtual correction of pneumencephalus.

CONCLUSION:

The technique provided a clear improvement in fusion of pre- and post-implantation scans, although the accuracy is difficult to evaluate.
Handle
https://boris-portal.unibe.ch/handle/20.500.12422/126110
Show full item
File(s)
FileFile TypeFormatSizeLicensePublisher/Copright statementContent
StieglitzEtAlNeurosurgery2014.pdftextAdobe PDF678.8 KBacceptedOpen
BORIS Portal
Bern Open Repository and Information System
Build: 27ad28 [15.10. 15:21]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo