  3. Phantom-based evaluation of yttrium-90 datasets using biograph vision quadra.
 

Phantom-based evaluation of yttrium-90 datasets using biograph vision quadra.

Options
  • Details
  • Files
BORIS DOI
10.48350/175767
Publisher DOI
10.1007/s00259-022-06074-3
PubMed ID
36504278
Description
PURPOSE

The image quality characteristics of two NEMA phantoms with yttrium-90 (90Y) were evaluated on a long axial field-of-view (AFOV) PET/CT. The purpose was to identify the optimized reconstruction setup for the imaging of patients with hepatocellular carcinoma after 90Y radioembolization.

METHODS

Two NEMA phantoms were used, where one had a 1:10 sphere to background activity concentration ratio and the second had cold background. Reconstruction parameters used are as follows: iterations 2 to 8, Gaussian filter 2- to 6-mm full-width-at-half-maximum, reconstruction matrices 440 × 440 and 220 × 220, high sensitivity (HS), and ultra-high sensitivity (UHS) modes. 50-, 40-, 30-, 20-, 10-, and 5-min acquisitions were reconstructed. The measurements included recovery coefficients (RC), signal-to-noise ratio (SNR), background variability, and lung error which measures the residual error in the corrections. Patient data were reconstructed with 20-, 10-, 5-, and 1-min time frames and evaluated in terms of SNR.

RESULTS

The RC for the hot phantom was 0.36, 0.45, 0.53, 0.63, 0.68, and 0.84 for the spheres with diameters of 10, 13, 17, 22, 28, and 37 mm, respectively, for UHS 2 iterations, a 220 × 220 matrix, and 50-min acquisition. The RC values did not differ with acquisition times down to 20 min. The SNR was the highest for 2 iterations, measured 11.7, 16.6, 17.6, 19.4, 21.9, and 27.7 while the background variability was the lowest (27.59, 27.08, 27.36, 26.44, 30.11, and 33.51%). The lung error was 18%. For the patient dataset, the SNR was 19%, 20%, 24%, and 31% higher for 2 iterations compared to 4 iterations for 20-, 10-, 5-, and 1-min time frames, respectively.

CONCLUSIONS

This study evaluates the NEMA image quality of a long AFOV PET/CT scanner with 90Y. It provides high RC for the smallest sphere compared to other standard AFOV scanners at shorter scan times. The maximum patient SNR was for 2 iterations, 20 min, while 5 min delivers images with acceptable SNR.
Date of Publication
2023-03
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Keyword(s)
Image quality NEMA Quadra Whole-body PET/CT Yttrium-90
Language(s)
en
Contributor(s)
Zeimpekis, Konstantinos
Universitätsklinik für Nuklearmedizin
Mercolli, Lorenzoorcid-logo
Universitätsklinik für Nuklearmedizin
Conti, Maurizio
Sari, Hasan
Prenosil, George
Universitätsklinik für Nuklearmedizin
Shi, Kuangyuorcid-logo
Universitätsklinik für Nuklearmedizin
Rominger, Axel Oliverorcid-logo
Universitätsklinik für Nuklearmedizin
Additional Credits
Universitätsklinik für Nuklearmedizin
Series
European journal of nuclear medicine and molecular imaging
Publisher
Springer
ISSN
1619-7089
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: dd892c [ 9.04. 8:30]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo