  3. Accelerated magnetic resonance diffusion tensor imaging of the median nerve using simultaneous multi-slice echo planar imaging with blipped CAIPIRINHA.
 

Accelerated magnetic resonance diffusion tensor imaging of the median nerve using simultaneous multi-slice echo planar imaging with blipped CAIPIRINHA.

Options
  • Details
  • Files
BORIS DOI
10.7892/boris.77239
Publisher DOI
10.1007/s00330-015-3985-8
PubMed ID
26373759
Description
PURPOSE

To investigate the feasibility of MR diffusion tensor imaging (DTI) of the median nerve using simultaneous multi-slice echo planar imaging (EPI) with blipped CAIPIRINHA.

MATERIALS AND METHODS

After federal ethics board approval, MR imaging of the median nerves of eight healthy volunteers (mean age, 29.4 years; range, 25-32) was performed at 3 T using a 16-channel hand/wrist coil. An EPI sequence (b-value, 1,000 s/mm(2); 20 gradient directions) was acquired without acceleration as well as with twofold and threefold slice acceleration. Fractional anisotropy (FA), mean diffusivity (MD) and quality of nerve tractography (number of tracks, average track length, track homogeneity, anatomical accuracy) were compared between the acquisitions using multivariate ANOVA and the Kruskal-Wallis test.

RESULTS

Acquisition time was 6:08 min for standard DTI, 3:38 min for twofold and 2:31 min for threefold acceleration. No differences were found regarding FA (standard DTI: 0.620 ± 0.058; twofold acceleration: 0.642 ± 0.058; threefold acceleration: 0.644 ± 0.061; p ≥ 0.217) and MD (standard DTI: 1.076 ± 0.080 mm(2)/s; twofold acceleration: 1.016 ± 0.123 mm(2)/s; threefold acceleration: 0.979 ± 0.153 mm(2)/s; p ≥ 0.074). Twofold acceleration yielded similar tractography quality compared to standard DTI (p > 0.05). With threefold acceleration, however, average track length and track homogeneity decreased (p = 0.004-0.021).

CONCLUSION

Accelerated DTI of the median nerve is feasible. Twofold acceleration yields similar results to standard DTI.

KEY POINTS

• Standard DTI of the median nerve is limited by its long acquisition time. • Simultaneous multi-slice acquisition is a new technique for accelerated DTI. • Accelerated DTI of the median nerve yields similar results to standard DTI.
Date of Publication
2016
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Keyword(s)
Diffusion tensor imaging
•
Diffusion tractography
•
Echo-planar imaging
•
Median nerve
•
Simultaneous multi-slice
Language(s)
en
Contributor(s)
Filli, Lukas
Piccirelli, Marco
Kenkel, David
Boss, Andreas
Manoliu, Andrei
Andreisek, Gustav
Bhat, Himanshu
Runge, Val Murray
Universitätsinstitut für Diagnostische, Interventionelle und Pädiatrische Radiologie
Guggenberger, Roman
Additional Credits
Universitätsinstitut für Diagnostische, Interventionelle und Pädiatrische Radiologie
Series
European radiology
Publisher
Springer
ISSN
0938-7994
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: dd892c [ 9.04. 8:30]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo