• LOGIN
    Login with username and password
Repository logo

BORIS Portal

Bern Open Repository and Information System

  • Publications
  • Theses
  • Research Data
  • Projects
  • Organizations
  • Researchers
  • More
  • Collections
  • Statistics
  • LOGIN
    Login with username and password
Repository logo
Unibern.ch
  1. Home
  2. Publications
  3. Implants with Original and Non-Original Abutment Connections
 

Implants with Original and Non-Original Abutment Connections

Options
  • Details
  • Files
BORIS DOI
10.7892/boris.11336
Publisher DOI
10.1111/j.1708-8208.2012.00479.x
PubMed ID
22804848
Description
AIM: To test in vitro the mechanical resistance, rotational misfit and failure mode of three original implant-abutment connections and to compare them to two connections between non-original abutments connected to one of the original implants. MATERIAL AND METHODS: Three different implants with small diameters (3.3 mm for Straumann Roxolid, 3.5 mm for Nobel Biocare Replace and Astra Tech Osseospeed TX) were connected with individualized titanium abutments. Twelve implants from each system were connected to their original abutments (Straumann CARES, Nobel Biocare Procera, Astra Tech Atlantis). Twenty-four Roxolid implants were connected with non-original abutments using CAD/CAM procedures from the other two manufacturers (12 Nobel Biocare Procera and 12 Astra Tech Atlantis). For the critical bending test, a Zwick/Roell 1475 machine and the Xpert Zwick/Roell software were used. RESULTS: The rotational misfit varied when comparing the different interfaces. The use of non-original grade V titanium abutments on Roxolid implants increased the force needed for deformation. The fracture mode was different with one of the original connections. CONCLUSIONS: Non-original abutments differ in design of the connecting surfaces and material and demonstrate higher rotational misfit. These differences may result in unexpected failure modes.
Date of Publication
2014-04
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Keyword(s)
abutment connection
•
abutments
•
CAD/CAM
•
dental implants
•
ISO 14801
•
non-original
•
original
•
titanium grade IV
•
titanium grade V
Language(s)
en
Contributor(s)
Gigandet, Michel
Gianni, Bigolin
Francisco, Faoro
Bürgin-Mumenthaler, Walter Bruno
Zahnmedizinische Kliniken, Forschung
Brägger, Urs
Zahnmedizinische Kliniken, Kronen- und Brückenprothetik
Zahnmedizinische Kliniken, Klinik für Rekonstruktive Zahnmedizin und Gerodontologie
Additional Credits
Zahnmedizinische Kliniken, Forschung
Zahnmedizinische Kliniken, Kronen- und Brückenprothetik
Series
Clinical implant dentistry and related research
Publisher
Blackwell
ISSN
1523-0899
Access(Rights)
restricted
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: dd892c [ 9.04. 8:30]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo