An alternative radiation-free method for assessing the accuracy of robotic computer-assisted implant surgery: a cross-sectional study.
Options
BORIS DOI
Publisher DOI
PubMed ID
41581748
Description
Objectives
This study aimed to assess the accuracy of implant placement performed using robotic computer-assisted implant surgery (r-CAIS) by comparing postoperative measurements obtained via intraoral scanning (IOS) with those derived from cone-beam computed tomography (CBCT).Methods
The study employed a cross-sectional design. All patients received a preoperative CBCT and IOS for surgical planning and registration. All the surgeries were performed using r-CAIS. Postoperative CBCT and IOS datasets were acquired for each patient and superimposed on the preoperative datasets to assess deviations between planned and actual implant positions. Paired t-tests were used to analyze differences between the two assessment modalities, with statistical significance defined as P < 0.05. Agreement between the two methods was evaluated using intra-class correlation coefficients (ICCs) and Bland-Altman analyses.Results
Thirty-four implants in twenty-seven patients were included. Global coronal, global apical, and angular deviations measured using the IOS method were 0.59 ± 0.23 mm, 0.66 ± 0.24 mm, and 1.28 ± 0.69°, respectively, compared with 0.59 ± 0.24 mm, 0.63 ± 0.25 mm, and 1.19 ± 0.70° measured by the CBCT method. No statistically significant differences were observed between the two modalities (P > 0.05, respectively). ICCs for global coronal, global apical, and angular deviations were 0.911, 0.886, and 0.856, respectively. Bland-Altman analyses further confirmed excellent agreement between the two methods.Conclusions
Both IOS and CBCT methods demonstrated comparable accuracy and good agreement in r-CAIS.Clinical Significance
Postoperative IOS method may serve as a reliable and radiation-free alternative to CBCT for assessing implant placement accuracy in r-CAIS in patients with a single missing tooth or two consecutive missing teeth. In future, multicenter randomized controlled trials with larger sample sizes are warranted to further validate these findings.
This study aimed to assess the accuracy of implant placement performed using robotic computer-assisted implant surgery (r-CAIS) by comparing postoperative measurements obtained via intraoral scanning (IOS) with those derived from cone-beam computed tomography (CBCT).Methods
The study employed a cross-sectional design. All patients received a preoperative CBCT and IOS for surgical planning and registration. All the surgeries were performed using r-CAIS. Postoperative CBCT and IOS datasets were acquired for each patient and superimposed on the preoperative datasets to assess deviations between planned and actual implant positions. Paired t-tests were used to analyze differences between the two assessment modalities, with statistical significance defined as P < 0.05. Agreement between the two methods was evaluated using intra-class correlation coefficients (ICCs) and Bland-Altman analyses.Results
Thirty-four implants in twenty-seven patients were included. Global coronal, global apical, and angular deviations measured using the IOS method were 0.59 ± 0.23 mm, 0.66 ± 0.24 mm, and 1.28 ± 0.69°, respectively, compared with 0.59 ± 0.24 mm, 0.63 ± 0.25 mm, and 1.19 ± 0.70° measured by the CBCT method. No statistically significant differences were observed between the two modalities (P > 0.05, respectively). ICCs for global coronal, global apical, and angular deviations were 0.911, 0.886, and 0.856, respectively. Bland-Altman analyses further confirmed excellent agreement between the two methods.Conclusions
Both IOS and CBCT methods demonstrated comparable accuracy and good agreement in r-CAIS.Clinical Significance
Postoperative IOS method may serve as a reliable and radiation-free alternative to CBCT for assessing implant placement accuracy in r-CAIS in patients with a single missing tooth or two consecutive missing teeth. In future, multicenter randomized controlled trials with larger sample sizes are warranted to further validate these findings.
Date of Publication
2026-04
Publication Type
Article
Subject(s)
Keyword(s)
Accuracy
•
Cone beam computed tomography
•
Intraoral scan
•
Robotic computer-assisted implant surgery
Language(s)
en
Contributor(s)
Zhu, Rongrong | |
He, Longlong | |
Zhou, Qin | |
Zhang, Meng | |
Chang, Xiaofeng |
Additional Credits
Series
Journal of Dentistry
Publisher
Elsevier
ISSN
1879-176X
0300-5712
Access(Rights)
open.access