High-accuracy patient-to-image registration for the facilitation of image-guided robotic microsurgery on the head
Options
Publisher DOI
PubMed ID
23340586
Description
Image-guided microsurgery requires accuracies an order of magnitude higher than today's navigation systems provide. A critical step toward the achievement of such low-error requirements is a highly accurate and verified patient-to-image registration. With the aim of reducing target registration error to a level that would facilitate the use of image-guided robotic microsurgery on the rigid anatomy of the head, we have developed a semiautomatic fiducial detection technique. Automatic force-controlled localization of fiducials on the patient is achieved through the implementation of a robotic-controlled tactile search within the head of a standard surgical screw. Precise detection of the corresponding fiducials in the image data is realized using an automated model-based matching algorithm on high-resolution, isometric cone beam CT images. Verification of the registration technique on phantoms demonstrated that through the elimination of user variability, clinically relevant target registration errors of approximately 0.1 mm could be achieved.
Date of Publication
2013
Publication Type
Article
Subject(s)
Language(s)
en
Contributor(s)
Series
IEEE transactions on biomedical engineering
Publisher
Institute of Electrical and Electronics Engineers IEEE
ISSN
0018-9294
Access(Rights)
metadata.only