Resonance Frequency Analysis (RFA) during early healing
Options
Description
Objectives: - to monitor resonance frequency analysis (RFA) in relation to the jawbone characteristics during the early phases of healing and incorporation of Straumann® dental implants with an SLA surface. Material and methods: 17 Straumann 4.1mm implants (10mm) and 7 Straumann 4.8mm implants (10mm) were installed and ISQ determined at baseline and after 1, 2, 3, 4, 5, 6, 8 and 12 weeks. Central bone cores were analyzed from the 4.1mm implants using micro CT for bone volume density (BVD) and bone trabecular connectivity (BTC). Results: Pocket probing depths ranged between 2-4mm and bleeding on probing between 5-20%. At baseline, BVD varied between 24 and 65% and BTC between 4.9 and 25.4 for the 4.1mm implants. Baseline ISQ varied between 55 and 74 with a mean of 61.4. No significant correlations were found between BVD or BTC and ISQ Values. For the 4.8mm diameter implants baseline ISQ values ranged from 57 – 70 with a mean of 63.3. Over the healing period ISQ values increased at 1 week and decreased after 2-3 weeks. After 4 weeks ISQ values, again increased slightly, no significant differences were noted over time. One implant (4.1mm) lost stability at 3 weeks. Its ISQ value had dropped from 68 to 45. However the latter value was determined after the clinical diagnosis of instability. Conclusion: ISQ values of 57 – 70 represented homeostasis and implant stability. However no predictive value for loosing implant stability can be attributed to RFA since decease occurred after the fact.