Publication:
Comparison of Hemodynamic Performance, Three-Dimensional Flow Fields, and Turbulence Levels for Three Different Heart Valves at Three Different Hemodynamic Conditions.

cris.virtual.author-orcid0000-0002-6062-9076
cris.virtualsource.author-orcid21a10c45-8d4f-4088-88a4-22a4c711cb11
cris.virtualsource.author-orcidfe58815c-ad76-46e4-912c-5be3fa73f92a
dc.contributor.authorFerrari, Lorenzo
dc.contributor.authorObrist, Dominik
dc.date.accessioned2024-11-07T11:14:48Z
dc.date.available2024-11-07T11:14:48Z
dc.date.issued2024-12
dc.description.abstractThe hemodynamic performance of different prosthetic heart valves is difficult to compare among studies due to a variety of test conditions and experimental techniques. Existing studies are typically limited to one family of valves (biological or mechanical) and testing conditions of 5l/min and often lack sufficient spatial resolution. To address these limitations, a pulse duplicator with a multi-view imaging system (Tomo-PIV) was employed to investigate the three-dimensional flow field in the aortic root of three different valves: a tri-leaflet mechanical heart valve (TRIFLO, Novostia), a bi-leaflet mechanical heart valve (On-X, Artivion), and a biological heart valve (Perimount, Edwards Lifesciences). The valves were tested at low (3 l/min), normal (5 l/min), and elevated (7 l/min) cardiac output under hypotensive (40/60mmHg), normotensive (80/120mmHg), and moderate hypertensive (105/170mmHg) pressure conditions, respectively. Compared to the Perimount, peak mean velocity was - 33%, - 24%, - 18% for the TRIFLO and - 32%, - 20%, - 11% for the On-X at low, moderate, and elevated , respectively. Corresponding peak values decreased by - 66%, - 57%, - 44% (TRIFLO) and - 60%, - 50%, - 36% (On-X). At low , was lower for Perimount (1.07cm) than for TRIFLO (1.47cm) and On-X (1.52cm), while it increased for elevated to 2.75cm (TRIFLO) and 2.16cm (Perimount and On-X). For all valves, increasing led to increased flow velocities, higher and higher levels of turbulence, and the spatial influence of the valve on the flow field in the ascending aorta was extended. peaked closer to the STJ than for TRIFLO and Perimount.
dc.description.numberOfPages12
dc.description.sponsorshipARTORG Center - Cardiovascular Engineering (CVE)
dc.description.sponsorshipARTORG Center - Cardiovascular Engineering (CVE)
dc.identifier.doi10.48620/76092
dc.identifier.pmid39287910
dc.identifier.publisherDOI10.1007/s10439-024-03584-z
dc.identifier.urihttps://boris-portal.unibe.ch/handle/20.500.12422/103417
dc.language.isoen
dc.publisherSpringer
dc.relation.ispartofAnnals of Biomedical Engineering
dc.relation.issn0090-6964
dc.subjectAortic valve replacement
dc.subjectBHV
dc.subjectBMHV
dc.subjectTRIFLO
dc.subjectTomo-PIV
dc.titleComparison of Hemodynamic Performance, Three-Dimensional Flow Fields, and Turbulence Levels for Three Different Heart Valves at Three Different Hemodynamic Conditions.
dc.typearticle
dspace.entity.typePublication
oaire.citation.endPage3207
oaire.citation.issue12
oaire.citation.startPage3196
oaire.citation.volume52
oairecerif.author.affiliationARTORG Center - Cardiovascular Engineering (CVE)
oairecerif.author.affiliationARTORG Center - Cardiovascular Engineering (CVE)
oairecerif.author.affiliation2ARTORG Center for Biomedical Engineering Research
unibe.contributor.rolecorresponding author
unibe.contributor.roleauthor
unibe.description.ispublishedpub
unibe.refereedtrue
unibe.subtype.articlejournal

Files

Original bundle
Now showing 1 - 1 of 1
Name:
s10439-024-03584-z.pdf
Size:
3.14 MB
Format:
Adobe Portable Document Format
File Type:
text
License:
https://creativecommons.org/licenses/by/4.0
Content:
published

Collections