  1. Home
  2. Publications
  3. Safety and feasibility of peri-device leakage closure after LAAO: an international, multicentre collaborative study.
 

Safety and feasibility of peri-device leakage closure after LAAO: an international, multicentre collaborative study.

Options
  • Details
  • Files
BORIS DOI
10.48350/163276
Publisher DOI
10.4244/EIJ-D-21-00286
PubMed ID
34219662
Description
BACKGROUND

Residual peri-device leakage (PDL) is frequent after left atrial appendage occlusion (LAAO). Little is known about management strategies, procedural aspects and outcomes of interventional PDL closure.

AIMS

The aim of this study was to assess the safety and feasibility of PDL closure after LAAO.

METHODS

Fifteen centres contributed data on baseline characteristics, in-hospital and follow-up outcomes of patients who underwent PDL closure after LAAO. Outcomes of interest included acute success and complication rates and long-term efficacy of the procedure.

RESULTS

A total of 95 patients were included and a cumulative number of 104 leaks were closed. The majority of PDLs were detected within 90 days (range 41-231). Detachable coils were the most frequent approach (42.3%), followed by the use of the AMPLATZER Vascular Plug II (29.8%) and the AMPLATZER Duct Occluder II (17.3%). Technical success was 100% with 94.2% of devices placed successfully within the first attempt. There were no major complications requiring surgical or transcatheter interventions. During follow-up (96 days [range 49-526]), persistent leaks were found in 18 patients (18.9%), yielding a functional success rate of 82.7%, although PDLs were significantly reduced in size (pre-leak sizemax: 6.1±3.6 mm vs post-leak sizemax: 2.5±1.3 mm, p<0.001). None of the patients had a leak >5 mm. Major adverse events during follow-up occurred in 5 patients (2 ischaemic strokes, 2 intracranial haemorrhages, and 1 major gastrointestinal bleeding).

CONCLUSIONS

Several interventional techniques have become available to achieve PDL closure. They are associated with high technical and functional success and low complication rates.
Date of Publication
2021-12-17
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Language(s)
en
Contributor(s)
Piayda, Kerstin
Sievert, Kolja
Della Rocca, Domenico G
Adeola, Oluwaseun G
Alkhouli, Mohamad
Yoo, David
Benito-González, Tomas
Cruz-González, Ignatio
Galea, Roberto
Universitätsklinik für Kardiologie
Skurk, Carsten
De Backer, Ole
Nielsen-Kudsk, Jens Erik
Grygier, Marek
Beaty, Elijah H
Newton, Jim
Pérez de Prado, Armando
Räber, Lorenz
Universitätsklinik für Kardiologie
Gibson, Douglas
Van Niekerk, Christoffel
Ellis, Christopher R
Horton, Rodney P
Natale, Andrea
Grundwald, Iris
Zeus, Tobias
Sievert, Horst
Additional Credits
Universitätsklinik für Kardiologie
Series
EuroIntervention
Publisher
Europa Digital & Publishing
ISSN
1774-024X
Access(Rights)
restricted
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: dd892c [ 9.04. 8:30]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo