  2. Publications
  3. Feasibility of Using EMG for Early Detection of the Facial Nerve During Robotic Direct Cochlear Access
 

Feasibility of Using EMG for Early Detection of the Facial Nerve During Robotic Direct Cochlear Access

Options
  • Details
  • Files
BORIS DOI
10.7892/boris.43575
Publisher DOI
10.1097/MAO.0000000000000187
PubMed ID
24492132
Description
HYPOTHESIS

Facial nerve monitoring can be used synchronous with a high-precision robotic tool as a functional warning to prevent of a collision of the drill bit with the facial nerve during direct cochlear access (DCA).

BACKGROUND

Minimally invasive direct cochlear access (DCA) aims to eliminate the need for a mastoidectomy by drilling a small tunnel through the facial recess to the cochlea with the aid of stereotactic tool guidance. Because the procedure is performed in a blind manner, structures such as the facial nerve are at risk. Neuromonitoring is a commonly used tool to help surgeons identify the facial nerve (FN) during routine surgical procedures in the mastoid. Recently, neuromonitoring technology was integrated into a commercially available drill system enabling real-time monitoring of the FN. The objective of this study was to determine if this drilling system could be used to warn of an impending collision with the FN during robot-assisted DCA.

MATERIALS AND METHODS

The sheep was chosen as a suitable model for this study because of its similarity to the human ear anatomy. The same surgical workflow applicable to human patients was performed in the animal model. Bone screws, serving as reference fiducials, were placed in the skull near the ear canal. The sheep head was imaged using a computed tomographic scanner and segmentation of FN, mastoid, and other relevant structures as well as planning of drilling trajectories was carried out using a dedicated software tool. During the actual procedure, a surgical drill system was connected to a nerve monitor and guided by a custom built robot system. As the planned trajectories were drilled, stimulation and EMG response signals were recorded. A postoperative analysis was achieved after each surgery to determine the actual drilled positions.

RESULTS

Using the calibrated pose synchronized with the EMG signals, the precise relationship between distance to FN and EMG with 3 different stimulation intensities could be determined for 11 different tunnels drilled in 3 different subjects.

CONCLUSION

From the results, it was determined that the current implementation of the neuromonitoring system lacks sensitivity and repeatability necessary to be used as a warning device in robotic DCA. We hypothesize that this is primarily because of the stimulation pattern achieved using a noninsulated drill as a stimulating probe. Further work is necessary to determine whether specific changes to the design can improve the sensitivity and specificity.
Date of Publication
2014-03
Publication Type
Article
Subject(s)
600 Technology > 630 Agriculture
600 Technology > 610 Medicine & health
Language(s)
en
Contributor(s)
Anso, Juanorcid-logo
ARTORG Center - Image Guided Therapy
Precht, Maria Christinaorcid-logo
Departement für klinische Veterinärmedizin (DKV)
Departement klinische Veterinärmedizin, Klinische Radiologie
Gerber, Nicolasorcid-logo
ARTORG Center - Artificial Hearing Research
Williamson, Tom
ARTORG Center - Image Guided Therapy
Gerber, Kateorcid-logo
ARTORG Center - Image Guided Therapy
Rösler, Kai Michael
Departement Klinische Forschung, Forschungsgruppe Neurologie
Caversaccio, Marco
Universitätsklinik für Hals-, Nasen- und Ohrenkrankheiten (HNO)
Weber, Stefan
ARTORG Center - Image Guided Therapy
Bell, Brett
ARTORG Center - Image Guided Therapy
Additional Credits
Universitätsklinik für Hals-, Nasen- und Ohrenkrankheiten (HNO)
ARTORG Center - Image Guided Therapy
Departement für klinische Veterinärmedizin (DKV)
ARTORG Center - Artificial Hearing Research
Departement Klinische Forschung, Forschungsgruppe Neurologie
Series
Otology & neurotology
Publisher
Lippincott Williams & Wilkins
ISSN
1531-7129
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: dd892c [ 9.04. 8:30]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo