  2. Publications
  3. TH-AB-201-05: Determining the Direction Distribution of the Primary Radiation for a Cyberknife-M6.
 

TH-AB-201-05: Determining the Direction Distribution of the Primary Radiation for a Cyberknife-M6.

Options
  • Details
Publisher DOI
10.1118/1.4958033
Description
PURPOSE

Radiation protection regulatory differentiates between primary and scatter radiation. Whereas for conventional clinical linear accelerators the solid angle for primary radiation is planar, the Cyberknife (Accuray Inc., Sunnyvale, CA) may point its beam in all spatial directions. In order to be able to judge on radiation protection calculations for a Cyberknife-M6 vault, the direction distribution for delivered plans was evaluated based on clinical experiences.

METHODS

The log-files of 121 delivered patient treatment plans were exported, divided into cranial and extra-cranial treatments and the delivered monitor units (MU) together with the corresponding beam directions were analyzed. This MU-weighted spatial distribution was then projected to a 9.5 × 5.9 × 3.9 m(3) vault, generating an "intensity map" using a binning of 50 × 50 cm(2) . The factor of direction (FOD) is reported as a fraction of the total applied MUs to the walls, ceiling and floor in the perspective of a patient lying in head-first-supine position on the couch. In this study, the term intFOD refers to the integral FOD and maxFOD refers to the maximal FOD for a single bin.

RESULTS

For all kind of treatments and collimators, intFOD and maxFOD for the wall behind the patient's head is 0.0. The intFOD for the floor varies between 0.65 and 0.74. For the ceiling, maxFOD is 0.002 and 0.0 for cranial and extra-cranial cases, respectively. The intFOD for the wall at the patient's feet, is 0.094 for cranial and 0.005 for extra-cranial cases. There is nearly no difference between the maxFOD of the right and left wall for cranial cases, whereas for extra-cranial cases these numbers differ by a factor of 1.75.

CONCLUSION

The varying FODs for the Cyberknife were quantified based on delivered treatment plans. These findings are helpful regarding the design of Cyberknife vaults.
Date of Publication
2016-06
Publication Type
Conference Item
Subject(s)
600 Technology > 610 Medicine & health
Keyword(s)
Collimators
•
Linear accelerators
•
Radiation treatment
•
Spatial dimensions
•
Therapeutics
Language(s)
en
Contributor(s)
Henzen, Dominikorcid-logo
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
Zanella do Amaral Campos, Pedro Paulo
Universitätsklinik für Intensivmedizin
Schmidhalter, Danielorcid-logo
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
Volken, Werner
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
Mackeprang, Paul-Henry
Universitätsklinik für Radio-Onkologie
Malthaner, Marco
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
Fix, Michaelorcid-logo
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
Manser, Peter
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
Additional Credits
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
Universitätsklinik für Intensivmedizin
Universitätsklinik für Radio-Onkologie
Series
Medical physics
Publisher
American Association of Physicists in Medicine AAPM
ISSN
0094-2405
Title of Event
AAPM Annual Meeting
Access(Rights)
metadata.only
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: dd892c [ 9.04. 8:30]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo