  3. Dose calculation of dynamic trajectory radiotherapy using Monte Carlo.
 

Dose calculation of dynamic trajectory radiotherapy using Monte Carlo.

Options
  • Details
  • Files
BORIS DOI
10.7892/boris.116291
Publisher DOI
10.1016/j.zemedi.2018.03.002
PubMed ID
29631759
Description
PURPOSE

Using volumetric modulated arc therapy (VMAT) delivery technique gantry position, multi-leaf collimator (MLC) as well as dose rate change dynamically during the application. However, additional components can be dynamically altered throughout the dose delivery such as the collimator or the couch. Thus, the degrees of freedom increase allowing almost arbitrary dynamic trajectories for the beam. While the dose delivery of such dynamic trajectories for linear accelerators is technically possible, there is currently no dose calculation and validation tool available. Thus, the aim of this work is to develop a dose calculation and verification tool for dynamic trajectories using Monte Carlo (MC) methods.

METHODS

The dose calculation for dynamic trajectories is implemented in the previously developed Swiss Monte Carlo Plan (SMCP). SMCP interfaces the treatment planning system Eclipse with a MC dose calculation algorithm and is already able to handle dynamic MLC and gantry rotations. Hence, the additional dynamic components, namely the collimator and the couch, are described similarly to the dynamic MLC by defining data pairs of positions of the dynamic component and the corresponding MU-fractions. For validation purposes, measurements are performed with the Delta4 phantom and film measurements using the developer mode on a TrueBeam linear accelerator. These measured dose distributions are then compared with the corresponding calculations using SMCP. First, simple academic cases applying one-dimensional movements are investigated and second, more complex dynamic trajectories with several simultaneously moving components are compared considering academic cases as well as a clinically motivated prostate case.

RESULTS

The dose calculation for dynamic trajectories is successfully implemented into SMCP. The comparisons between the measured and calculated dose distributions for the simple as well as for the more complex situations show an agreement which is generally within 3% of the maximum dose or 3mm. The required computation time for the dose calculation remains the same when the additional dynamic moving components are included.

CONCLUSION

The results obtained for the dose comparisons for simple and complex situations suggest that the extended SMCP is an accurate dose calculation and efficient verification tool for dynamic trajectory radiotherapy. This work was supported by Varian Medical Systems.
Date of Publication
2019-02
Publication Type
Article
Keyword(s)
Dose calculation Dynamic trajectories Monte Carlo Verification
Language(s)
en
Contributor(s)
Manser, Peter
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
Frauchiger, Daniel
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
Frei, Daniel
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
Volken, Werner
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
Terribilini, Dario
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
Fix, Michaelorcid-logo
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
Additional Credits
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik
Series
Zeitschrift für medizinische Physik
Publisher
Elsevier
ISSN
1876-4436
Access(Rights)
restricted
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: dd892c [ 9.04. 8:30]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo