Transapical transcatheter aortic valve implantation using the second-generation self-expanding Symetis ACURATE TA valve
Options
BORIS DOI
Publisher DOI
PubMed ID
25298363
Description
Transapical transcatheter aortic valve implantation (TA-TAVI) is the recognized first choice surgical TAVI access. Expansion of this well-established treatment modality with subsequent broader patient inclusion has accelerated development of second-generation TA-TAVI devices. The Swiss ACURATE TA Symetis valve allows for excellent anatomical positioning, resulting in a very low incidence of paravalvular leaks. The self-expanding stent features an hourglass shape to wedge the native aortic valve annulus. A specially designed delivery system facilitates controlled release aided by tactile operator feedback. The ACURATE TA valve made of three native porcine non-coronary leaflets has received CE approval in September 2011. Since then, this valve is the third most frequently implanted TAVI device with over 1200 implants in Europe and South America. Results from the Symetis ACURATE TA™ Valve Implantation ('SAVI') Registry showed a procedural success rate of 98.0% and a survival rate of 93.2% at 30 days. This presentation provides technical considerations and detailed procedural aspects of device implantation.
Date of Publication
2014
Publication Type
Article
Subject(s)
Keyword(s)
Nitinol
•
Results
•
Symetis device
•
Technique
•
Transapical aortic valve implantation
Language(s)
en
Series
Multimedia Manual of Cardiothoracic Surgery
Publisher
Oxford University Press
ISSN
1813-9175
Access(Rights)