  3. Method-comparison study between a watch-like sensor and a cuff-based device for 24-h ambulatory blood pressure monitoring.
 

Method-comparison study between a watch-like sensor and a cuff-based device for 24-h ambulatory blood pressure monitoring.

Options
  • Details
  • Files
BORIS DOI
10.48350/191181
Publisher DOI
10.1038/s41598-023-33205-z
PubMed ID
37061569
Description
The use of 24-h ambulatory blood pressure monitoring (ABPM) has been continuously increasing over the last decades. However, cuff-based devices may cause discomfort, particularly at night, leading to potentially non-representative blood pressure (BP) values. We investigated the feasibility of a cuff-less BP monitoring solution in 67 subjects undergoing conventional 24-h ABPM. A watch-like optical sensor was attached at the upper arm or wrist at the contralateral side of the cuff. Systolic (SBP) and diastolic BP (DBP) values were estimated from the measured optical signals by pulse wave analysis. Average 24-h, daytime and nighttime BP values were compared between the conventional monitor and the cuff-less sensor. The differences between both methods-expressed as mean ± standard deviation (95% limits of agreement)-were of - 1.8 ± 6.2 mmHg (- 13.9, 10.3) on SBP and - 2.3 ± 5.4 mmHg (- 13.0, 8.3) on DBP for 24-h averages, of - 1.5 ± 6.6 mmHg (- 14.4, 11.4) on SBP and - 1.8 ± 5.9 mmHg (- 13.4, 9.9) on DBP for daytime averages, and of 0.4 ± 7.5 mmHg (- 14.4, 15.1) on SBP and - 1.3 ± 6.8 mmHg (- 14.7, 12.0) on DBP for nighttime averages. These results encouragingly suggest that cuff-less 24-h ABPM may soon become a clinical possibility.
Date of Publication
2023-04-15
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Language(s)
en
Contributor(s)
Proença, Martin
Ambühl, Jeremias
Bonnier, Guillaume
Meister, Théo Arthur Perceval
Universitätsklinik für Kardiologie
Valentin, Jérémy Pierre Patrick
Universitätsklinik für Kardiologie
Soria, Rodrigo
Ferrario, Damien
Lemay, Mathieu
Rexhaj, Emrush
Universitätsklinik für Kardiologie
Additional Credits
Universitätsklinik für Kardiologie
Series
Scientific reports
Publisher
Springer Nature
ISSN
2045-2322
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: dd892c [ 9.04. 8:30]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo