[Seasonal fluctuations and influence of nutrition on macular pigment density]
Options
BORIS DOI
Date of Publication
2006
Publication Type
Article
Division/Institute
Author
Jahn, C | |
Brinkmann, C | |
Mössner, A | |
Wüstemeyer, H | |
Schnurrbusch, U |
Series
Ophthalmologe
ISSN or ISBN (if monograph)
0941-293X
Publisher
Springer-Verlag
Language
English
Publisher DOI
PubMed ID
16341834
Description
PURPOSE: We have previously reported on measuring macular pigment density (MPD) with a scanning laser ophthalmoscope (HRA, Heidelberg Engineering, Heidelberg, Germany). This study war undertaken to evaluate the variation of MPD over a period of 1 year in healthy subjects. METHOD: We used autofluorescence images recorded with a HRA to evaluate MPD with a 2 degrees circle centered on the fovea. Healthy subjects were included in the study and MPD measurements were repeated every 2 months over a period of 1 year. RESULTS: We included a total of 30 healthy subjects aged 19-34 years (mean: 23+/-2 years). Mean MPD at time point 1 was 0.215+/-0.056 density units (DU), at time point 2 0.235+/-0.051 DU, at time point 3 0.218+/-0.055 DU, at time point 4 0.228+/-0.057 DU, at time point 5 0.225+/-0.053 DU, and at time point 6 0.203+/-0.050 DU. The statistical analysis revealed no significant variation of MPD over the follow-up period of 1 year. CONCLUSION: This study demonstrates that MPD shows no variation over a period of 1 year in healthy subjects.
File(s)
File | File Type | Format | Size | License | Publisher/Copright statement | Content | |
---|---|---|---|---|---|---|---|
347_2005_Article_1282.pdf | text | Adobe PDF | 721.43 KB | publisher | published |