Effects of an Automated External Defibrillator With Additional Video Instructions on the Quality of Cardiopulmonary Resuscitation

Options
  • Details
  • Files
BORIS DOI
10.48350/154274
Publisher DOI
10.3389/fmed.2021.640721
PubMed ID
33816528
Description
Aim of the Study: The aim was to compare cardiopulmonary resuscitation (CPR) quality of an automated external defibrillator (AED) with and without additional video instruction during basic life support (BLS) by laypersons.

Methods: First-year medical students were randomized either to an AED with audio only or audio with additional video instructions during CPR. Each student performed 4 min of single-rescuer chest compression only BLS on a manikin (Ambu Man C, Ballerup, Denmark) using the AED. The primary outcome was the effective compression ratio during this scenario. This combined parameter was used to evaluate the quality of chest compressions by multiplying compressions with correct depth, correct hand position, and complete decompression by flow time. Secondary outcomes were percentages of incomplete decompression and hand position, mean compression rate, time-related parameters, and subjective assessments.

Results: Effective compression ratio did not differ between study groups in the overall sample (p = 0.337) or in students with (p = 0.953) or without AED experience (p = 0.278). Additional video instruction led to a higher percentage of incorrect decompressions (p = 0.014). No significant differences could be detected in time-related resuscitation parameters. An additional video was subjectively rated as more supporting (p = 0.001).

Conclusions: Audio–video instructions did not significantly improve resuscitation quality in these laypersons despite that it was felt more supportive. An additional video to the verbal AED prompts might lead to cognitive overload. Therefore, future studies might target the influence of the video content and the potential benefits of video instructions in specific populations.
Date of Publication
2021-03-17
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Language(s)
en
Contributor(s)
Ettl, Florian
Fischer, Eva
Losert, Heidrun
Stumpf, Dominik
Ristl, Robin
Ruetzler, Kurt
Greif, Robertorcid-logo
Universitätsklinik für Anästhesiologie und Schmerztherapie
Fischer, Henrik
Additional Credits
Universitätsklinik für Anästhesiologie und Schmerztherapie
Series
Frontiers in medicine
Publisher
Frontiers
ISSN
2296-858X
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: dd892c [ 9.04. 8:30]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo