  3. Al2O3:C optically stimulated luminescence dosimeters (OSLDs) for ultra-high dose rate proton dosimetry.
 

Al2O3:C optically stimulated luminescence dosimeters (OSLDs) for ultra-high dose rate proton dosimetry.

Options
  • Details
  • Files
BORIS DOI
10.48350/160333
Publisher DOI
10.1088/1361-6560/abe554
PubMed ID
33571973
Description
The response of Al2O3:C optically stimulated luminescence detectors (OSLDs) was investigated in a 250 MeV pencil proton beam. The OSLD response was mapped for a wide range of average dose rates up to 9000 Gy s-1, corresponding to a ∼150 kGy s-1instantaneous dose rate in each pulse. Two setups for ultra-high dose rate (FLASH) experiments are presented, which enable OSLDs or biological samples to be irradiated in either water-filled vials or cylinders. The OSLDs were found to be dose rate independent for all dose rates, with an average deviation <1% relative to the nominal dose for average dose rates of (1-1000) Gy s-1when irradiated in the two setups. A third setup for irradiations in a 9000 Gy s-1pencil beam is presented, where OSLDs are distributed in a 3 × 4 grid. Calculations of the signal averaging of the beam over the OSLDs were in agreement with the measured response at 9000 Gy s-1. Furthermore, a new method was presented to extract the beam spot size of narrow pencil beams, which is in agreement within a standard deviation with results derived from radiochromic films. The Al2O3:C OSLDs were found applicable to support radiobiological experiments in proton beams at ultra-high dose rates.
Date of Publication
2021-04-15
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Keyword(s)
Al2O3:C FLASH optically stimulated luminescence proton dosimetry ultra-high dose rate
Language(s)
en
Contributor(s)
Christensen, Jeppe Brage
Togno, Michele
Nesteruk, Konrad Pawel
Psoroulas, Serena
Meer, David
Weber, Damien Charles
Universitätsklinik für Radio-Onkologie
Lomax, Tony
Yukihara, Eduardo G
Safai, Sairos
Additional Credits
Universitätsklinik für Radio-Onkologie
Series
Physics in medicine and biology
Publisher
Institute of Physics Publishing IOP
ISSN
0031-9155
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: dd892c [ 9.04. 8:30]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo