Publication:
The Shelhigh No-React bovine internal mammary artery: a questionable alternative conduit in coronary bypass surgery?

cris.virtualsource.author-orcid63ad244b-000b-4e0e-be7f-80bddc233642
cris.virtualsource.author-orcid04943b8f-3791-4bf5-ac35-7f5cdd30f14a
cris.virtualsource.author-orcid4383feca-67b2-4dc5-8cca-815e432ad487
cris.virtualsource.author-orcidb7097efb-27db-46f7-a2c5-0d762b820361
cris.virtualsource.author-orcidfb99e59d-56d8-4e1d-87a2-10152d09d1ca
datacite.rightsopen.access
dc.contributor.authorEnglberger, Lars
dc.contributor.authorNoti, Jasmin
dc.contributor.authorImmer, Franz Ferdinand
dc.contributor.authorStalder, Mario
dc.contributor.authorEckstein, Friedrich Stefan
dc.contributor.authorCarrel, Thierry
dc.date.accessioned2024-10-13T18:20:20Z
dc.date.available2024-10-13T18:20:20Z
dc.date.issued2008
dc.description.abstractBACKGROUND: Increasing age and comorbidities among patients undergoing coronary artery bypass surgery (CABG) stimulates the exhaustive research for alternative grafts. No-React treatment should render the tissue resistant against degeneration and reduce early inflammatory response. The aim of the present study was an invasive assessment of the patency of No-React bovine internal mammary artery (NRIMA grafts) used as bypass conduit in CABG surgery. PATIENTS AND METHODS: Nineteen NRIMA grafts were used in 17 patients (2.9%) out of a total of 572 patients undergoing CABG surgery within a 12-month period. All intraoperative data were assessed and in-hospital outcome was analysed. Follow-up examination was performed 7.0+/-4.0 months after initial surgery, including clinical status and coronary angiography to assess patency of the NRIMA grafts. RESULTS: Average perioperative flow of all NRIMA grafts was 71+/-60 ml/min. One patient died in hospital due to a multi-organ failure. Four patients refused invasive assessment. Follow-up was complete in 12 patients with overall 13 NRIMA grafts. Nine NRIMA grafts (69.2%) were used for the right coronary system, two NRIMA grafts (15.4%) on the LAD and two on the circumflex artery. Graft patency was 23.1% and was independent of the intraoperative flow measurement. CONCLUSIONS: NRIMA grafts show a very low patency and cannot be recommended as coronary bypass graft conduits. Patency was independent of the perioperative flow, assessed by Doppler ultrasound. Because of this unsatisfying observation, this type of graft should be utilised as a last resource conduit and used only to revascularise less important target vessels, such as the end branches of the right coronary artery.
dc.description.numberOfPages3
dc.description.sponsorshipUniversitätsklinik für Herz- und Gefässchirurgie
dc.identifier.doi10.7892/boris.28281
dc.identifier.isi000253752500019
dc.identifier.pmid18083039
dc.identifier.publisherDOI10.1016/j.ejcts.2007.11.006
dc.identifier.urihttps://boris-portal.unibe.ch/handle/20.500.12422/101674
dc.language.isoen
dc.publisherElsevier Science B.V.
dc.publisher.placeOxford
dc.relation.isbn18083039
dc.relation.ispartofEuropean journal of cardio-thoracic surgery
dc.relation.issn1010-7940
dc.relation.organizationDCD5A442BADFE17DE0405C82790C4DE2
dc.titleThe Shelhigh No-React bovine internal mammary artery: a questionable alternative conduit in coronary bypass surgery?
dc.typearticle
dspace.entity.typePublication
dspace.file.typetext
oaire.citation.endPage4
oaire.citation.issue2
oaire.citation.startPage222
oaire.citation.volume33
oairecerif.author.affiliationUniversitätsklinik für Herz- und Gefässchirurgie
oairecerif.author.affiliationUniversitätsklinik für Herz- und Gefässchirurgie
oairecerif.author.affiliationUniversitätsklinik für Herz- und Gefässchirurgie
oairecerif.author.affiliationUniversitätsklinik für Herz- und Gefässchirurgie
oairecerif.author.affiliationUniversitätsklinik für Herz- und Gefässchirurgie
unibe.contributor.rolecreator
unibe.contributor.rolecreator
unibe.contributor.rolecreator
unibe.contributor.rolecreator
unibe.contributor.rolecreator
unibe.contributor.rolecreator
unibe.date.licenseChanged2019-11-20 22:49:05
unibe.description.ispublishedpub
unibe.eprints.legacyId28281
unibe.journal.abbrevTitleEUR J CARDIO-THORAC SURG
unibe.refereedtrue
unibe.subtype.articlejournal

Files

Original bundle
Now showing 1 - 1 of 1
Name:
33-2-222.pdf
Size:
68.12 KB
Format:
Adobe Portable Document Format
File Type:
text
License:
publisher
Content:
published

Collections