  3. Use of straight and curved 3-dimensional titanium miniplates for fracture fixation at the mandibular angle

Use of straight and curved 3-dimensional titanium miniplates for fracture fixation at the mandibular angle

Details
Files
DOI
10.7892/boris.23762
Official URL
http://www.sciencedirect.com/science/article/pii/S0278239107003382
Publisher DOI
10.1016/j.joms.2007.03.013
PubMed ID
17719394
Abstract
PURPOSE: The aim of this follow-up study was to evaluate the clinical usefulness of a new type of 3-dimensional (3D) miniplate for open reduction and monocortical fixation of mandibular angle fractures. PATIENTS AND METHODS: In 20 consecutive patients, noncomminuted mandibular angle fractures were treated with open reduction and fixation using a 2 mm 3D miniplate system in a transoral approach. All patients were systematically monitored until 6 months postoperatively. Among the outcome parameters recorded were infection, hardware failure, wound dehiscence, and sensory disturbance of the inferior alveolar nerve. RESULTS: The mean operation time from incision to wound closure was 65 minutes. Two patients had a mucosal wound dehiscence with no consequences. None developed an infection requiring a plate removal. All but 2 patients had normal sensory function 3 months after surgery. Plate fracture occurred in one patient in whom a preceding surgical removal of the third molar had been the reason for the mandibular fracture. In the absence of clinical symptoms, the patient declined plate removal. On final follow-up, fracture healing was considered clinically complete in all patients. CONCLUSIONS: The 3D plating system described here is suitable for fixation of simple mandibular angle fractures and is an easy-to-use alternative to conventional miniplates. The system may be contraindicated in patients in whom insufficient interfragmentary bone contact causes minor stability of the fracture.
Date Issued
2007
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Language(s)
en
Author(s)
Zix, Jürgen  
Universitätsklinik für Schädel-, Kiefer- und Gesichtschirurgie  
Lieger, Olivier  
Universitätsklinik für Schädel-, Kiefer- und Gesichtschirurgie  
Iizuka, Tateyuki  
Universitätsklinik für Schädel-, Kiefer- und Gesichtschirurgie  
Additional Credits
Universitätsklinik für Schädel-, Kiefer- und Gesichtschirurgie  
Journal
Journal of oral and maxillofacial surgery
Publisher
Elsevier
ISSN
0278-2391
ISBN
17719394
Access(Rights)
restricted
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: 0eaa7c [ 7.08. 11:06]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • BORIS Portal & Open Science
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo
Repository logo COAR Notify