  3. Technical feasibility of constant-load and high-intensity interval training for cardiopulmonary conditioning using a re-engineered dynamic leg press
 

Technical feasibility of constant-load and high-intensity interval training for cardiopulmonary conditioning using a re-engineered dynamic leg press

Options
  • Details
  • Files
BORIS DOI
10.7892/boris.137973
Publisher DOI
10.1186/s42490-019-0025-9
Description
Background
Leg-press devices are one of the most widely used training tools for musculoskeletal strengthening of the lower-limbs, and have demonstrated important cardiopulmonary benefits for healthy and patient populations. Further engineering development was done on a dynamic leg-press for work-rate estimation by integrating force and motion sensors, power calculation and a visual feedback system for volitional work-rate control. This study aimed to assess the feasibility of the enhanced dynamic leg press for cardiopulmonary exercise training in constant-load training and high-intensity interval training. Five healthy participants aged 31.0±3.9 years (mean ± standard deviation) performed two cardiopulmonary training sessions: constant-load training and high-intensity interval training. Participants carried out the training sessions at a work rate that corresponds to their first ventilatory threshold for constant-load training, and their second ventilatory threshold for high-intensity interval training.

Results
All participants tolerated both training protocols, and could complete the training sessions with no complications. Substantial cardiopulmonary responses were observed. The difference between mean oxygen uptake and target oxygen uptake was 0.07±0.34 L/min (103 ±17%) during constant-load training, and 0.35±0.66 L/min (113 ±27%) during high-intensity interval training. The difference between mean heart rate and target heart rate was −7±19 bpm (94 ±15%) during constant-load training, and 4.2±16 bpm (103 ±12%) during high-intensity interval training.

Conclusions
The enhanced dynamic leg press was found to be feasible for cardiopulmonary exercise training, and for exercise prescription for different training programmes based on the ventilatory thresholds.
Date of Publication
2019-10-03
Publication Type
Article
Subject(s)
500 - Science::570 - Life sciences; biology
600 - Technology::610 - Medicine & health
600 - Technology::620 - Engineering
Language(s)
en
Contributor(s)
Chrif, Farouk
ARTORG Center for Biomedical Engineering Research
Nef, Tobiasorcid-logo
ARTORG Center - Gerontechnology and Rehabilitation
Hunt, Kenneth J
Additional Credits
ARTORG Center for Biomedical Engineering Research
ARTORG Center - Gerontechnology and Rehabilitation
Series
BMC biomedical engineering
Publisher
BioMed Central
ISSN
2524-4426
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: ae9592 [15.12. 16:43]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo