Comparison of tendency-oriented perimetry and dynamic strategy in octopus perimetry as a screening tool in a clinical setting: a prospective study
Options
Publisher DOI
PubMed ID
17458786
Description
OBJECTIVE: The aim of this study was to compare the results of tendency-oriented perimetry (TOP) and a dynamic strategy in octopus perimetry as screening methods in clinical practice. DESIGN: A prospective single centre observational case series was performed. PARTICIPANTS AND METHODS: In a newly opened general ophthalmologic practice 89 consecutive patients (171 eyes) with a clinical indication for octopus static perimetry testing (ocular hypertension or suspicious optic nerve cupping) were examined prospectively with TOP and a dynamic strategy. The visual fields were graded by 3 masked observers as normal, borderline or abnormal without any further clinical information. RESULTS: 83% eyes showed the same result for both strategies. In 14% there was a small difference (with one visual field being abnormal or normal, the other being borderline). In only 2.9% of the eyes (5 cases) was there a contradictory result. In 4 out of 5 cases the dynamic visual field was abnormal and TOP was normal. 4 of these cases came back for a second examination. In all 4 the follow-up examination showed a normal second dynamic visual field. CONCLUSIONS: Octopus static perimetry using a TOP strategy is a fast, patient-friendly and very reliable screening tool for the general ophthalmological practice. We found no false-negative results in our series.
Date of Publication
2007
Publication Type
Article
Language(s)
en
Contributor(s)
Scherrer, M | |
Helbig, H | |
Johann Auf der Heide, K | |
Sutter, F K |
Additional Credits
Series
Klinische Monatsblätter für Augenheilkunde
Publisher
Ferdinand Enke Verlag
ISSN
0023-2165
ISBN
17458786
Access(Rights)
metadata.only