Cervical soft tissue imaging using a mobile CBCT scanner with a flat panel detector in comparison with corresponding CT and MRI data sets
Options
Publisher DOI
PubMed ID
17900944
Description
OBJECTIVE: The aim of this study was to evaluate soft tissue image quality of a mobile cone-beam computed tomography (CBCT) scanner with an integrated flat-panel detector. STUDY DESIGN: Eight fresh human cadavers were used in this study. For evaluation of soft tissue visualization, CBCT data sets and corresponding computed tomography (CT) and magnetic resonance imaging (MRI) data sets were acquired. Evaluation was performed with the help of 10 defined cervical anatomical structures. RESULTS: The statistical analysis of the scoring results of 3 examiners revealed the CBCT images to be of inferior quality regarding the visualization of most of the predefined structures. Visualization without a significant difference was found regarding the demarcation of the vertebral bodies and the pyramidal cartilages, the arteriosclerosis of the carotids (compared with CT), and the laryngeal skeleton (compared with MRI). Regarding arteriosclerosis of the carotids compared with MRI, CBCT proved to be superior. CONCLUSIONS: The integration of a flat-panel detector improves soft tissue visualization using a mobile CBCT scanner.
Date of Publication
2007
Publication Type
Article
Language(s)
en
Contributor(s)
Heiland, Max | |
Pohlenz, Philipp | |
Blessmann, Marco | |
Habermann, Christian R | |
Begemann, Philipp C | |
Schmidgunst, Christian | |
Blake, Felix A S | |
Püschel, Klaus | |
Schmelzle, Rainer | |
Schulze, Dirk |
Additional Credits
Series
Oral surgery, oral medicine, oral pathology, oral radiology and endodontology
Publisher
Elsevier
ISSN
1079-2104
ISBN
17900944
Access(Rights)
metadata.only