A system for 3-D reconstruction of a patient-specific surface model from calibrated X-ray images
Publisher DOI
Abstract
This paper presents a system for 3-D reconstruction of a patient-specific surface model from calibrated X-ray images. Our system requires two X-ray images of a patient with one acquired from the anterior-posterior direction and the other from the axial direction. A custom-designed cage is utilized in our system to calibrate both images. Starting from bone contours that are interactively identified from the X-ray images, our system constructs a patient-specific surface model of the proximal femur based on a statistical model based 2D/3D reconstruction algorithm. In this paper, we present the design and validation of the system with 25 bones. An average reconstruction error of 0.95 mm was observed.
Date Issued
2009
Publication Type
Book Section
Language(s)
en
Editor(s)
Westwood, James D | |
Westwood, Susan W | |
Haluck, Randy S | |
Hoffman, Helene M | |
Mogel, Greg T | |
Phillips, Roger | |
Robb, Richard A | |
Vosburgh, Kirby G |
Additional Credits
Publisher
IOS Press
ISSN
0926-9630
ISBN
978-1-60750-402-3
Book Title
Access(Rights)
metadata.only