  3. Next generation viscoelasticity assays in cardiothoracic surgery: Feasibility of the TEG6s system.
 

Next generation viscoelasticity assays in cardiothoracic surgery: Feasibility of the TEG6s system.

Options
  • Details
  • Files
BORIS DOI
10.7892/boris.123136
Publisher DOI
10.1371/journal.pone.0209360
PubMed ID
30571781
Description
BACKGROUND

Viscoelastic near-patient assays of global hemostasis have been found useful and cost-effective in perioperative settings. Shortcomings of current systems include substantial laboratory intensity, user-dependent reproducibility, relatively large sample volumes, sensitivity to ambient vibration and limited comparability between techniques and devices. The aim of this study was to assess feasibility of a new, resonance-based viscoelastic whole blood methodology (TEG6s) in cardiac surgery with cardiopulmonary bypass (CPB) and to compare the parameters this system produces with the ROTEM delta system and standard coagulation tests.

METHODS

In a prospective evaluation study, twenty-three consecutive cardiac surgery patients underwent hemostasis management according to current guidelines, using the ROTEM delta system and standard coagulation tests. Blood samples were collected prior to CPB before anesthetic induction (pre-CPB), during CPB on rewarming (CPB), and 10 minutes after heparin reversal with protamine (post-CPB). ROTEM and standard coagulation test results were compared with TEG6s parameters, which were concurrently determined using its multi-channel microfluidic cartridge system.

RESULTS

TEG6s provided quantifiable results pre-CPB and post-CPB, but only R (clotting time) of CKH (kaolin with heparinase) was measurable during CPB (full heparinization). Spearman's correlation coefficient (rs) was 0.78 for fibrinogen levels and MA CFF (functional fibrinogen). Correlation of several TEG6s parameters was good (0.77 to 0.91) with MCF FIBTEM, and poor (<0.56) with prothrombin time and activated partial thromboplastin time (<0.44). Rs with platelet count was moderate (0.70, MA CK; 0.73, MA CRT). Accuracy of MA CFF for detection of fibrinogen deficiency < 1.5 g/L was high (ROC-AUC 0.93).

CONCLUSIONS

The TEG6s system, which is based on resonance viscoelastic methodology, appears to be feasible for POC hemostasis assessment in cardiac surgery. Its correlations with standard coagulation parameters are quite similar to those of ROTEM and there is good diagnostic accuracy for fibrinogen levels lower than 1.5 g/L. During full heparinization, TEG6s testing is limited to R measurement. Larger studies are needed to assess superiority over other POC systems.
Date of Publication
2018-12-20
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Language(s)
en
Contributor(s)
Erdoes, Gabor
Universitätsklinik für Anästhesiologie und Schmerztherapie
Schloer, Hannes
Eberle, Balthasarorcid-logo
Universitätsklinik für Anästhesiologie und Schmerztherapie
Nagler, Michael
Department for BioMedical Research, Forschungsgruppe Hämatologie (Erwachsene)
Universitätsklinik für Hämatologie und Hämatologisches Zentrallabor
Additional Credits
Universitätsklinik für Anästhesiologie und Schmerztherapie
Department for BioMedical Research, Forschungsgruppe Hämatologie (Erwachsene)
Series
PLoS ONE
Publisher
Public Library of Science
ISSN
1932-6203
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: dd892c [ 9.04. 8:30]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo