  3. Benchmarking Monte-Carlo dose calculation for MLC CyberKnife treatments.

Benchmarking Monte-Carlo dose calculation for MLC CyberKnife treatments.

Details
Files
DOI
10.7892/boris.133501
Publisher DOI
10.1186/s13014-019-1370-5
PubMed ID
31533746
Abstract
BACKGROUND

Vendor-independent Monte Carlo (MC) dose calculation (IDC) for patient-specific quality assurance of multi-leaf collimator (MLC) based CyberKnife treatments is used to benchmark and validate the commercial MC dose calculation engine for MLC based treatments built into the CyberKnife treatment planning system (Precision MC).

METHODS

The benchmark included dose profiles in water in 15 mm depth and depth dose curves of rectangular MLC shaped fields ranging from 7.6 mm × 7.7 mm to 115.0 mm  × 100.1 mm, which were compared between IDC, Precision MC and measurements in terms of dose difference and distance to agreement. Dose distributions of three phantom cases and seven clinical lung cases were calculated using both IDC and Precision MC. The lung PTVs ranged from 14 cm3 to 93 cm3. Quantitative comparison of these dose distributions was performed using dose-volume parameters and 3D gamma analysis with 2% global dose difference and 1 mm distance criteria and a global 10% dose threshold. Time to calculate dose distributions was recorded and efficiency was assessed.

RESULTS

Absolute dose profiles in 15 mm depth in water showed agreement between Precision MC and IDC within 3.1% or 1 mm. Depth dose curves agreed within 2.3% / 1 mm. For the phantom and clinical lung cases, mean PTV doses differed from - 1.0 to + 2.3% between IDC and Precision MC and gamma passing rates were > =98.1% for all multiple beam treatment plans. For the lung cases, lung V20 agreed within ±1.5%. Calculation times ranged from 2.2 min (for 39 cm3 PTV at 1.0 × 1.0 × 2.5 mm3 native CT resolution) to 8.1 min (93 cm3 at 1.1 × 1.1 × 1.0 mm3), at 2% uncertainty for Precision MC for the 7 examined lung cases and 4-6 h for IDC, which, however, is not optimized for efficiency but used as a gold standard for accuracy.

CONCLUSIONS

Both accuracy and efficiency of Precision MC in the context of MLC based planning for the CyberKnife M6 system were benchmarked against MC based IDC framework. Precision MC is used in clinical practice at our institute.
Date Issued
2019-09-18
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Subjects
Benchmarking CyberKnife Dose calculation Monte Carlo QA TPS
Language(s)
en
Author(s)
Mackeprang, Paul-Henry  
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik  
Vuong, D
Volken, Werner  
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik  
Henzen, Dominik  
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik  
Schmidhalter, Daniel  
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik  
Malthaner, Marco  
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik  
Müller, Silvan Andreas  
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik  
Frei, Daniel  
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik  
Kilby, W
Aebersold, Daniel Matthias  
Universitätsklinik für Radio-Onkologie  
Universitätsklinik für Radio-Onkologie  
Fix, Michael  
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik  
Manser, Peter  
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik  
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik  
Additional Credits
Universitätsklinik für Radio-Onkologie, Medizinische Strahlenphysik  
Universitätsklinik für Radio-Onkologie  
Journal
Radiation oncology
Publisher
BioMed Central
ISSN
1748-717X
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: 0eaa7c [ 7.08. 11:06]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • BORIS Portal & Open Science
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo
Repository logo COAR Notify