  3. Comparison of Fixation Techniques for Acetabular Fractures Involving the Anterior Column with Disruption of the Quadrilateral Plate: A Biomechanical Study.

Comparison of Fixation Techniques for Acetabular Fractures Involving the Anterior Column with Disruption of the Quadrilateral Plate: A Biomechanical Study.

Details
Files
DOI
10.7892/boris.131527
Publisher DOI
10.2106/JBJS.17.00295
PubMed ID
29916932
Abstract
BACKGROUND

In elderly patients who have sustained an acetabular fracture involving disruption of the quadrilateral plate (QLP), postoperative loading of the joint beyond the level of partial weight-bearing can result in medial redisplacement of the QLP. The purpose of this biomechanical study was to compare the performances of 4 different fixation constructs intended to prevent medial redisplacement of the QLP.

METHODS

Anterior column posterior hemitransverse (ACPHT) fractures with disruption of the QLP were created on synthetic hemipelves (fourth-generation Sawbones models) and subsequently stabilized with (1) a 12-hole plate bridging the QLP (Group 1), (2) the plate with added periarticular screws along the QLP (Group 2), (3) the plate combined with an infrapectineal buttress plate (Group 3), or (4) the plate with the added periarticular screws as well as the buttress plate (Group 4). The point of load application on the acetabulum was defined to be the same as the point of application of maximum vertical hip contact force during normal walking. Loads were applied to simulate either partial weight-bearing (20 cycles, from 35 to 350 N) or inadvertent supraphysiologic loads (linearly increasing loads until the onset of failure, defined as fragment displacement of >3 mm). A universal testing machine was synchronized with a digital image correlation system to optically track redisplacement at the QLP. The level of significance was set at p < 0.05.

RESULTS

During experimental simulation of partial weight-bearing, maximum fracture step openings never exceeded 2 mm. During simulation of inadvertent supraphysiologic load, the median load to failure was higher (p < 0.05) in Group 2 (962 N; range, 798 to 1,000 N) and Group 4 (985 N; range, 887 to 1,000 N) compared with Group 1 (445 N; range, 377 to 583 N) and Group 3 (671 N; range, 447 to 720 N).

CONCLUSIONS

All 4 fixation constructs performed in an acceptable manner on testing with simulated partial weight-bearing. Only additional periarticular screws along the QLP increased the fixation strength.

CLINICAL RELEVANCE

Redisplacement of the QLP resulting in an incongruency of the hip joint has been associated with poor long-term outcomes. Within the constraints of this study, periarticular long screws were superior to infrapectineal buttress plates in preventing medial redisplacement of the QLP.
Date Issued
2018-06-20
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Language(s)
en
Author(s)
May, Christian
Egloff, Mike
Butscher, Andre
Keel, Marius  
Universitätsklinik für Orthopädische Chirurgie und Traumatologie  
Aebi, This
Siebenrock, Klaus-Arno  
Universitätsklinik für Orthopädische Chirurgie und Traumatologie  
Bastian, Johannes Dominik  
Universitätsklinik für Orthopädische Chirurgie und Traumatologie  
Additional Credits
Universitätsklinik für Orthopädische Chirurgie und Traumatologie  
Journal
The journal of bone and joint surgery
Publisher
Wolters Kluwer
ISSN
1535-1386
Access(Rights)
restricted
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: 0eaa7c [ 7.08. 11:06]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • BORIS Portal & Open Science
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo
Repository logo COAR Notify