The effect of standard and low-modulus cement augmentation on the stiffness, strength, and endplate pressure distribution in vertebroplasty
Publisher DOI
PubMed ID
22170449
Abstract
Vertebroplasty restores stiffness and strength of fractured vertebral bodies, but alters their stress transfer. This unwanted effect may be reduced by using more compliant cements. However, systematic experimental comparison of structural properties between standard and low-modulus augmentation needs to be done. This study investigated how standard and low-modulus cement augmentation affects apparent stiffness, strength, and endplate pressure distribution of vertebral body sections.
Date Issued
2012
Publication Type
Article
Language(s)
en
Author(s)
Kinzl, Michael | |
Boger, Andreas | |
Pahr, Dieter H |
Journal
European spine journal
Publisher
Springer
ISSN
0940-6719
Access(Rights)