  1. Home
  2. Publications
  3. Incidence and Predictors of Successful Lead Extraction With Short Active Mechanical Extraction Sheaths Only.

Incidence and Predictors of Successful Lead Extraction With Short Active Mechanical Extraction Sheaths Only.

Details
Files
DOI
10.48620/97697
Publisher DOI
10.1111/jce.70342
PubMed ID
41964517
Abstract
Background And Aims
Transvenous lead extraction (TLE) is increasingly performed for malfunction or infection of cardiac implantable electronic devices (CIED). Active mechanical rotational sheaths facilitate extraction and are available in short and long sizes, but it is unclear how many patients short active mechanical extraction sheaths (SAMES) are sufficient for TLE, and in which patients they may facilitate TLE. This study aimed to evaluate the incidence and predictors of successful lead extraction using only SAMES.Methods
We retrospectively analyzed patients who underwent TLE at the University Hospital Zurich from 2013 until 2023 regarding the use of short or long active mechanical extraction sheaths, as well as clinical and radiological success and complications.Results
A total of 568 TLEs were performed from January 2013 until December 2023. In 55 cases (9.7%), TLE was successful using only SAMES. Independent predictors for successful lead extraction using only SAMES in multivariate analysis were shorter lead dwell time (OR: 0.98, 95% CI: 0.97-0.99, p < 0.001) and use of anticoagulation (OR: 0.44, 95% CI: 0.19-1.00, p = 0.05). Complications occurred at a lower rate when only SAMES were used (All complications 12.7% vs. 24.2%, minor complications 9.1% vs. 18.7%, perioperative death 0% vs. 1.7%).Conclusion
In 10% of patients undergoing TLE, SAMES are sufficient for successful lead extraction. SAMES may be a sensible choice as the primary extraction tool in patients with chronic anticoagulation and shorter lead dwell time, as these factors predict a higher likelihood of success in TLE with SAMES only.
Date Issued
2026-06
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Subjects
active mechanical sheaths
•
complications
•
lead extraction
•
outcome
Language(s)
en
Author(s)
Wiederkehr, Caroline
Kuster, Noah
Bebié, Michelle
Steffel, Jan
Haeberlin, Andreas  
Clinic of Cardiology  
Breitenstein, Alexander
Hofer, Daniel
Additional Credits
Clinic of Cardiology  
Journal
Journal of Cardiovascular Electrophysiology
Publisher
Wiley
ISSN
1540-8167
1045-3873
Access(Rights)
open.access
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: 0eaa7c [ 7.08. 11:06]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • BORIS Portal & Open Science
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo
Repository logo COAR Notify