  3. In vivo demonstration of reflection artifact reduction in photoacoustic imaging using synthetic aperture photoacoustic-guided focused ultrasound (PAFUSion)

In vivo demonstration of reflection artifact reduction in photoacoustic imaging using synthetic aperture photoacoustic-guided focused ultrasound (PAFUSion)

Details
Files
DOI
10.7892/boris.96958
Publisher DOI
10.1364/BOE.7.002955
Abstract
Reflection artifacts caused by acoustic inhomogeneities are a critical problem in epi-mode biomedical photoacoustic imaging. High light fluence beneath the probe results in photoacoustic transients, which propagate into the tissue and reflect back from echogenic structures. These reflection artifacts cause problems in image interpretation and significantly impact the contrast and imaging depth. We recently proposed a method called PAFUSion (Photoacoustic-guided focused ultrasound) to identify such reflection artifacts in photoacoustic imaging. In its initial version, PAFUSion mimics the inward-travelling wavefield from small blood vessel-like PA sources by applying ultrasound pulses focused towards these sources, and thus provides a way to identify the resulting reflection artifacts. In this work, we demonstrate reduction of reflection artifacts in phantoms and in vivo measurements on human volunteers. In view of the spatially distributed PA sources that are found in clinical applications, we implemented an improved version of PAFUSion where photoacoustic signals are backpropagated to imitate the inward travelling wavefield and thus the reflection artifacts. The backpropagation is performed in a synthetic way based on the pulse-echo acquisitions after transmission on each single element of the transducer array. The results provide a direct confirmation that reflection artifacts are prominent in clinical epi-photoacoustic imaging, and that PAFUSion can strongly reduce these artifacts to improve deep-tissue photoacoustic imaging.
Date Issued
2016
Publication Type
Article
Subject(s)
600 Technology > 620 Engineering
500 Science
500 Science > 530 Physics
Language(s)
en
Author(s)
Singh, Mithun Kuniyil Ajith
Jaeger, Michael  
Institut für angewandte Physik (IAP)  
Frenz, Martin  
Institut für angewandte Physik (IAP)  
Steenbergen, Wiendelt
Additional Credits
Institut für angewandte Physik (IAP)  
Journal
Biomedical optics express
Publisher
Optical Society of America
ISSN
2156-7085
Access(Rights)
restricted
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: 0eaa7c [ 7.08. 11:06]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • BORIS Portal & Open Science
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo
Repository logo COAR Notify