  1. Home
  2. Publications
  3. Validation of 3D-reconstructed computed tomography images using OsiriX® software for pre-transcatheter aortic valve implantation aortic annulus sizing.

Validation of 3D-reconstructed computed tomography images using OsiriX® software for pre-transcatheter aortic valve implantation aortic annulus sizing.

Details
Files
DOI
10.7892/boris.101639
Publisher DOI
10.1093/icvts/ivx119
PubMed ID
28486661
Abstract
OBJECTIVES

We report validation of OsiriX -an image processing freeware-to measure multi-slice computed tomography-derived annulus diameters for preprocedural transcatheter aortic valve implantation planning.

METHODS

A total of 137 patients (82 ± 6.5 years, 42.3% male, logistic EuroSCORE 24.1 ± 14.2%) with severe aortic stenosis at high surgical risk underwent transcatheter aortic valve implantation assessment: transoesophageal echocardiography, angiography and multi-slice computed tomography. Retrospectively, 3D multi-slice computed tomography reconstructions were generated using OsiriX and the reliability evaluated regarding inter- and intraobserver variability, intermodality correlation and estimation of the clinical impact on transcatheter aortic valve implantation sizing.

RESULTS

Reliability of the novel OsiriX software was high with an interobserver mean difference of 0.6 ± 1.4 mm and intraclass correlation of absolute agreement of 0.84 (95% confidence interval 0.74-0.90). The intermodality accuracy between OsiriX measurements and conventional 2D computed tomography reconstructions, transoesophageal echocardiography and angiography revealed significantly larger sizing with OsiriX, with a mean difference to 2D computed tomography of 0.4 ± 2.2 mm, which would have changed valve sizing in 38% of patients. In 28%, a larger size would have been chosen, and this correlated highly with the occurrence of postoperative severe aortic regurgitation ( P  < 0.001).

CONCLUSIONS

While OsiriX measurements are an accurate and reproducible assessment of the aortic annulus, there are distinct and clinically relevant differences in aortic annulus dimensions between OsiriX measurements and previously standard imaging modalities. Sizing with OsiriX resulted in a larger perimeter compared with conventional 2D imaging. Careful assessment of valve size will take into account multiple imaging modalities.
Date Issued
2017-08-01
Publication Type
Article
Subject(s)
600 Technology > 610 Medicine & health
Subjects
3D imaging Aortic annulus dimensions TAVI prosthesis sizing
Language(s)
en
Author(s)
Gumsheimer, Marlena
Stortecky, Stefan  
Universitätsklinik für Kardiologie  
Gahl, Brigitta  
Universitätsklinik für Herz- und Gefässchirurgie  
Langhammer, Bettina  
Universitätsklinik für Herz- und Gefässchirurgie  
Carrel, Thierry  
Universitätsklinik für Herz- und Gefässchirurgie  
Büllesfeld, Lutz  
Universitätsklinik für Kardiologie  
Huber, Christoph  
Universitätsklinik für Herz- und Gefässchirurgie  
Most, Henriette  
Universitätsklinik für Herz- und Gefässchirurgie  
Additional Credits
Universitätsklinik für Herz- und Gefässchirurgie  
Universitätsklinik für Kardiologie  
Journal
Interactive cardiovascular and thoracic surgery
Publisher
Oxford University Press
ISSN
1569-9293
Access(Rights)
restricted
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: da6d1d [14.09. 16:14]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • BORIS Portal & Open Science
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo
Repository logo COAR Notify