  3. Cup Implant Planning Based on 2-D/3-D Radiographic Pelvis Reconstruction-First Clinical Results.

Cup Implant Planning Based on 2-D/3-D Radiographic Pelvis Reconstruction-First Clinical Results.

Details
Files
DOI
10.7892/boris.75252
Publisher DOI
10.1109/TBME.2015.2441378
PubMed ID
26057528
Abstract
GOAL:

In the following, we will present a newly developed X-ray calibration phantom and its integration for 2-D/3-D pelvis reconstruction and subsequent automatic cup planning. Two different planning strategies were applied and evaluated with clinical data.
METHODS:

Two different cup planning methods were investigated: The first planning strategy is based on a combined pelvis and cup statistical atlas. Thereby, the pelvis part of the combined atlas is matched to the reconstructed pelvis model, resulting in an optimized cup planning. The second planning strategy analyzes the morphology of the reconstructed pelvis model to determine the best fitting cup implant.
RESULTS:

The first planning strategy was compared to 3-D CT-based planning. Digitally reconstructed radiographs of THA patients with differently severe pathologies were used to evaluate the accuracy of predicting the cup size and position. Within a discrepancy of one cup size, the size was correctly identified in 100% of the cases for Crowe type I datasets and in 77.8% of the cases for Crowe type II, III, and IV datasets. The second planning strategy was analyzed with respect to the eventually implanted cup size. In seven patients, the estimated cup diameter was correct within one cup size, while the estimation for the remaining five patients differed by two cup sizes.
CONCLUSION:

While both planning strategies showed the same prediction rate with a discrepancy of one cup size (87.5%), the prediction of the exact cup size was increased for the statistical atlas-based strategy (56%) in contrast to the anatomically driven approach (37.5%).
SIGNIFICANCE:

The proposed approach demonstrated the clinical validity of using 2-D/3-D reconstruction technique for cup planning.
Date Issued
2015
Publication Type
Article
Subject(s)
500 Science > 570 Life sciences; biology
600 Technology > 610 Medicine & health
Language(s)
en
Author(s)
Schumann, Steffen  
Institut für chirurgische Technologien und Biomechanik (ISTB)  
Sato, Yoshinobu
Nakanishi, Yuki
Yokota, Futoshi
Takao, Masaki
Sugano, Nobuhiko
Zheng, Guoyan  orcid-logo
Institut für chirurgische Technologien und Biomechanik (ISTB)  
Additional Credits
Institut für chirurgische Technologien und Biomechanik (ISTB)  
Journal
IEEE transactions on biomedical engineering
Publisher
Institute of Electrical and Electronics Engineers IEEE
ISSN
0018-9294
Access(Rights)
restricted
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: da6d1d [14.09. 16:14]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • BORIS Portal & Open Science
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo
Repository logo COAR Notify