  3. In vivo performance of a laser fluorescence device for the approximal detection of caries in permanent molars

In vivo performance of a laser fluorescence device for the approximal detection of caries in permanent molars

Details
Publisher DOI
10.1016/j.jdent.2010.09.001
PubMed ID
20837089
Abstract
OBJECTIVES: The aim of this randomised clinical trial was to investigate if a laser fluorescence device is able to discriminate between sound and carious approximal sites and between enamel and dentinal lesions, as well as to find appropriate cut-off values. METHODS: One hundred and seventeen sound or uncavitated carious sites in permanent molars were visually and radiographically examined, then either opened or not, after which their laser fluorescence was measured. Forty-three lesions were opened, the caries removed and the clinically identified caries depths were registered in addition to the radiographical scoring. Seventy-four sites were radiographically deemed sound or had enamel caries and were not opened. Here, the radiographical scorings were registered. RESULTS: Taking the radiographic scoring as gold standard for all investigated approximal sites, sound sites (D(0), n=40) showed significantly lower laser fluorescence measurements than carious sites (D(1-4), n=77) (Mann-Whitney test, P<0.025) suggesting a cut-off at 7 (sensitivity=0.68, specificity=0.7). Comparing measurements of D(0-2) (n=74) and D(3,4) (n=43), the results were also different by a statistically significant amount (P<0.025) and the cut-off calculated to be 16 (sensitivity=0.6, specificity=0.84). A fair positive correlation between laser fluorescence values and radiographical scoring was found (rho=+0.47, P<0.01). Analysing the 43 opened lesions with their clinically found lesion depths as gold standard, there was a fair positive correlation to the laser fluorescence values (rho=+0.34, P=0.03) and a moderately strong correlation to the radiographic scoring (rho=+0.67, P<0.01). CONCLUSION: The device may be an adjunct tool in the approximal detection of caries along with established procedures.
Date Issued
2010
Publication Type
Article
Language(s)
en
Author(s)
Huth, K C
Lussi, Adrian  
Zahnmedizinische Kliniken, Klinik für Zahnerhaltung  
Gygax, Markus  
Zahnmedizinische Kliniken, Klinik für Zahnerhaltung  
Thum, M
Crispin, A
Paschos, E
Hickel, R
Neuhaus, Klaus  orcid-logo
Zahnmedizinische Kliniken, Klinik für Zahnerhaltung  
Additional Credits
Zahnmedizinische Kliniken, Klinik für Zahnerhaltung  
Journal
Journal of dentistry
Publisher
Elsevier Science
ISSN
0300-5712
Access(Rights)
metadata.only
Show full item
BORIS Portal
Bern Open Repository and Information System
Build: 24f0a9 [ 4.09. 8:55]
Explore
  • Projects
  • Funding
  • Publications
  • Research Data
  • Organizations
  • Researchers
  • Audiovisual Material
  • Software & other digital items
  • Events
More
  • About BORIS Portal
  • BORIS Portal & Open Science
  • Send Feedback
  • Cookie settings
  • Service Policy
Follow us on
  • Mastodon
  • YouTube
  • LinkedIn
UniBe logo
Repository logo COAR Notify